Retiro De Equipo (Recall) de Philips Ingenuity Core CT System

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Philips Medical Systems (USA).

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    22786
  • Fecha de inicio del evento
    2018-04-09
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    The manufacturer has become aware of a potential problem with the ingenuity core 128 family, that may have been infected with a virus before installation. the virus may encrypt files in a way that:, " prevents system access, or, " a patient undergoing a scan or recently scanned whose results have not been reviewed may require a ct rescan.
  • Acción
    Software to be upgraded

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    NZMMDSA