Retiro De Equipo (Recall) de Philips Pinnacle Radiation Treatment Planning System

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Philips Radiation Oncology Systems.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    17325
  • Fecha de inicio del evento
    2014-09-09
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    The issue occurs under certain specific conditions where the user modifies a contour in a density-overridden rol (region of interest with density set to a specific value) in a computed plan. dose is subsequently invalidated and the user may recompute the dose. in this case, the density can be replaced by the ct number from the ct to density table when recomputed resulting in an incorrect dose, and the monitor units (mu) may then be much higher than originally planned. the user may continue, and save the plan with incorrect dose. when the plan is reopened, the density is reported as the correct value but the dose remains incorrect based on the prior incorrect value for density.
  • Acción
    Software to be upgraded