Retiro De Equipo (Recall) de Raystation RaySearch treatment planning system

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por RaySearch Laboratories AB.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    16238
  • Fecha de inicio del evento
    2014-02-17
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: CARSL Consulting, 50B Puketapu Road, Taradale, Napier 4112
  • Causa
    The rotation component of the rigid registration computed prior to the deformable registration is not accounted for in fusion views when mapping structures (regions or points of interest) or when deforming dose. since the error affects all applications of the hybrid deformation, its effect on the above functions is displayed in the user interface. at that time the user would be able to view the result and evaluate the quality of the deformation and identify any potential errors in the results.
  • Acción
    Software to be upgraded

Device

  • Modelo / Serial
    Model: , Affected: , Software version: Versions: 2.5, 3.0, 3.5 and 4.0
  • Manufacturer

Manufacturer