Retiro De Equipo (Recall) de Roche Cobas Sysmex SP-10/ SP-1000 and associated transport systems (AlphaN, HST and XN-9000/3000)

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Roche Diagnostics GmbH.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    19796
  • Fecha de inicio del evento
    2016-02-27
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Roche Diagnostics (NZ) Ltd, 15 Rakino Way, Mt Wellington, AUCKLAND 1644
  • Causa
    There have been reports overseas that sample tubes may not be correctly placed back in the rack tube transport (rtt) after the de-capping process. this may result in open tubes being dropped within the cobas p 512 pre-analytical system, contaminating the system and potentially other samples. a false triggering or detection of the “ready” signal from the lifting gripper in the decapper causes the system to register that the motor motion is finished. the system then continues with the next process step and opens the gripper prematurely., this issue may lead to contamination of the instrument and other samples. the risk to operator of contamination is less than remote. however, the contamination of other samples may lead to erroneous sample results with possible inappropriate diagnosis/treatment.
  • Acción
    Software to be upgraded