Retiro De Equipo (Recall) de Siemens - _Magnetom Permanent

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Siemens Shenzhen Magnetic Resonance Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    19605
  • Fecha de inicio del evento
    2015-12-23
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Causa
    The manufacture confirms that in some cases the quench lines for some superconducting magnets may have not been properly installed and wants to alert the users to check if the quench lines for their systems have been installed under safe conditions. siemens in new zealand will inspect all quench lines installed by siemens., quench lines are installed to vent helium gas in the event of a quench. (quenching refers to the events that occur when the liquid cryogens that cool the magnet coils boil off rapidly, which results in helium escaping very rapidly from the cryogen bath) in the rare event of a quench an improperly installed quench line may cause the helium gas to be blown directly into the magnet room or other areas.
  • Acción
    Product to be modified

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    NZMMDSA