Retiro De Equipo (Recall) de Siemens Artis Zeego, Connector Robotic Stand

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Siemens AG.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    16373
  • Fecha de inicio del evento
    2014-03-12
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Causa
    Manufacturer confirms that in the course of continuous product monitoring inthe field, it has been discovered that in four individual cases the crimping of motor connectors a1 o a3 was incorrectly carried out. in these four cases incorrectly executed crimping in which the crimping tool was used on the wrong point of the contact led to substandard contacting.
  • Acción
    Product to be modified

Device

  • Modelo / Serial
    Model: , Affected: serial numbers of the motor connectors lower than 946295