Retiro De Equipo (Recall) de Siemens AXIOM- AX

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Siemens AG.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    19968
  • Fecha de inicio del evento
    2016-04-05
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Causa
    In artis zeego systems, angulations in the vicinity of the c-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. b) in artis systems with a100plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the x-ray tube) can result in the failure of a module in the high-voltage generator., follow up: the manufacturer has recently become aware of an additional affected system that we were not initially aware of. the additional affected customer is auckland hospital.
  • Acción
    Software to be upgraded