Retiro De Equipo (Recall) de Siemens Digital Linear Accelerator ARTISTE, ONCOR and PRIMUS

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Siemens AG.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    16786
  • Fecha de inicio del evento
    2014-06-04
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Causa
    Siemens became aware of an incident where a patient was pinched between the moving gantry and the tabletop during an auto-sequenced treatment. in this case, an automatic gantry movement during an auto-sequenced treatment led to the collision because:, * the auto-sequenced treatment has been created including beams with table angles., and, * no dry run has been performed., and, * the therapist did not monitor the patient during treatment delivery.
  • Acción
    Manufacturer to issue advice regarding use

Device

  • Modelo / Serial
    Model: , Affected:
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    NZMMDSA