Retiro De Equipo (Recall) de Stryker Surgical Light, spring arm

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Stryker Communications Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    13592
  • Fecha de inicio del evento
    2012-10-04
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Stryker New Zealand Ltd, 515 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
  • Causa
    Stryker has become aware that there is a small likelihood of incomplete engagement of the spring arm circlip following service or replacement of the spring arm. this has the potential to cause the spring arm to fall along with the surgical lights/monitors/navigation attachment mounted to the end of the spring arm.
  • Acción
    Product to be modified

Device

  • Modelo / Serial
    Model: Catalogue Numbers: 0682001270, 0682001274, 0682000226, 0682400059I, Affected: Selected lots