Retiro De Equipo (Recall) de Terumo Advanced Perfusion System 1

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Terumo Cardiovascular Systems Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    20499
  • Fecha de inicio del evento
    2016-08-15
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Terumo Australia Pty Ltd (NZ), c/- Healthcare Logistics, 56 Carrington Road, Mount Albert, Auckland 1025
  • Causa
    Reports have been received of issues that manifest themselves as power or communications failures in the system network. terumo's investigation revealed an internal cable with an inadequate crimp was installed in a subset of the total population of terumo system 1 units. the poor crimp in the cable can create a loss or degradation of internal power on one side of the system's base.
  • Acción
    Instructions for use to be updated