Retiro De Equipo (Recall) de The Binding Site Group Human IgM kit for use on SPA Plus

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por The Binding Site Group Ltd.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    15787
  • Fecha de inicio del evento
    2013-12-03
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: In Vitro Technologies, 26-30 Vestey Drive, Mount Wellington, Panmure, Auckland
  • Causa
    Manufacturer investigations identified that the affected product may be subject to antigen excess effect; even though it's true value was below the claimed level of 56.4g/l. the antigen excess capacity of this product has been reviewed and has been confirmed as correct. investigations have suggested that the tested sample contained predominantly monoclonal igm and the manufacturer believes that this particular type of igm reacted aberrantly with the antibody and subsequently yielded the spurious result.
  • Acción
    Instructions for use to be updated