Retiro De Equipo (Recall) de Therapeutic plasma exchange system

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Gambro Lundia AB.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    22901
  • Fecha de inicio del evento
    2018-05-07
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Baxter Healthcare Ltd, 33 Vestey Drive, Mt Wellington, AUCKLAND
  • Causa
    Manufacturer has identified that all prismaflex control units exhibit a failure mode with the pump module electronics. the failure mode may result in a voltage out of range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. manufacturer will be releasing new firmware that will prevent the malfunction from occurring.
  • Acción
    Software to be upgraded

Device

  • Modelo / Serial
    Model: 955052, 114870, 114489, 107493, 113082 and 113874, Affected:
  • Manufacturer

Manufacturer