Retiro De Equipo (Recall) de Zimmer Trabecular Metal Reverse Shoulder System

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Zimmer Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    15670
  • Fecha de inicio del evento
    2013-11-20
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Zimmer Biomet New Zealand Company, 210 Khyber Pass Road, Grafton, Auckland
  • Causa
    Manufacturer launced the zimmer trabecular metal reverse shoulder system in 2006, with over 30,000 sold since launch. during the last 7 years, zimmer has received 90 complaints related to the glenosphere disassociating from the base plate or difficulty assembling the glenosphere to the base plate. this represents a complaint rate of approximately 0.27% since the product was introduced. based on the investigation of these complaints, zimmer determined that additional instructions specific to the assembly of the glenosphere to the base plate were necessary in order to emphasize important techniques. as a result, zimmer updated the applicable surgical technique and also released an instructional video specific to the assembly of the glenosphere. zimmer are providing instructional video specific to the assembly of the glenosphere.
  • Acción
    Instructions for use to be updated

Device

  • Modelo / Serial
    Model: , Affected: 00-4349-036-11, 00-4349-038-11 and 00-4349-040-11
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    NZMMDSA