Retiro De Equipo (Recall) de Zimmer Trilogy Acetabular System Shells, without Holes

Según New Zealand Medicines and Medical Devices Safety Authority, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en New Zealand que fue producido por Zimmer Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    13667
  • Fecha de inicio del evento
    2012-11-15
  • País del evento
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Zimmer Biomet New Zealand Company, 210 Khyber Pass Road, Grafton, Auckland
  • Causa
    The manufacturer has determined that approximately 3% of trilogy acetabular shells without holes manufactured since march 2009 were slightly below the lower range of the specification (as low as 28% vs. 30% porosity specification). pore size and pore volume are two key characteristics of implant coatings.
  • Acción
    Manufacturer to issue advice regarding use

Device

  • Modelo / Serial
    Model: 80 models, Affected: Manufactured since March 2009
  • Manufacturer

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Comentario del fabricante
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    NZMMDSA