Notificaciones De Seguridad De Campo acerca de Beacon® Tip Royal Flush and others

Según Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), este evento ( notificaciones de seguridad de campo ) involucró a un dispositivo médico en Slovenia que fue producido por Cook Medical Europe.

¿Qué es esto?

Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • Fecha
    2016-04-21
  • País del evento
  • Fuente del evento
    AMPMDRS
  • Notas / Alertas
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notas adicionales en la data
  • Causa
    Cook medical is initiating a voluntary recall of the products listed above. we have identified an increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. our preliminary investigation into this matter has identified that environmental conditions, such as storage temperature, humidity, and the use of vaporized hydrogen peroxide (vhp) for whole-room decontamination within healthcare facilities, may be contributing to the occurrence. we also recognize that there may be other undetermined contributors to this issue and continue to investigate. .

Device

Manufacturer