Notificaciones De Seguridad De Campo acerca de Ceiling Supply Unit klinoPORT

Según Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), este evento ( notificaciones de seguridad de campo ) involucró a un dispositivo médico en Slovenia que fue producido por TRUMPF Medizin Systeme GmbH + Co. KG.

¿Qué es esto?

Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • Fecha
    2015-06-24
  • País del evento
  • Fuente del evento
    AMPMDRS
  • Notas / Alertas
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notas adicionales en la data
  • Causa
    Trumpf medical was informed of an event where the motorised height-adjustable arm of a klinoport ceiling supply unit suddenly and abruptly fell to its lowest position. the cause analysis revealed that the welded connections of the drive did not comply with the specified condition and therefore led to breakage of these connections. it is currently not to be ruled out that further drives within the same serial numbers range may have the same fault. if further cases should occur, the risk of severe injury to persons cannot be ruled out. .

Manufacturer