Notificaciones De Seguridad De Campo acerca de cobas u 601 / cobas u 701

Según Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), este evento ( notificaciones de seguridad de campo ) involucró a un dispositivo médico en Slovenia que fue producido por Roche Diagnostics.

¿Qué es esto?

Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • Fecha
    2017-12-21
  • País del evento
  • Fuente del evento
    AMPMDRS
  • Notas / Alertas
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notas adicionales en la data
  • Causa
    With this notification we would like to inform you about the possible sample mismatch within one rack. the analyzer may not transport the rack correctly, so that tubes might be detected twice. as a consequence, the same tube barcode label is read twice and the last tube barcode label is interpreted as the rack id. hence, a sample mismatch is possible between samples processed on the same rack, because the tubes are pipetted in the order given by the rack position and the test result assigned to the wrong sample id. the sample id read twice is thus listed in duplicate in the sample order list and test result list. the orders/test results have different rack ids. the consecutive sample id’s are also affected, since they are listed only once. as a consequence, they have an incorrect rack position displayed. .