Notificaciones De Seguridad De Campo acerca de Conformable GORE® TAG® Thoracic Endoprosthesis

Según Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), este evento ( notificaciones de seguridad de campo ) involucró a un dispositivo médico en Slovenia que fue producido por W. L. Gore & Associates, Inc..

¿Qué es esto?

Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • Fecha
    2017-10-23
  • País del evento
  • Fuente del evento
    AMPMDRS
  • Notas / Alertas
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notas adicionales en la data
  • Causa
    • since december 2016, gore has received four similar reports of incomplete and / or partial deployments of the conformable tag® device. in each event, the physician observed that half of the conformable tag® device deployed and half remained constrained to the delivery catheter. each of these events occurred during an off-label procedure, but it is unclear at this time how this may have affected the outcomes. engineering evaluations for two returned devices indicate that one partial deployment was the result of an incorrect deployment line stitch pattern and another was the result of deployment line damage of unknown origin. the other two devices have not been returned to gore for evaluation. • there were two serious adverse health consequences and one death reported: – one patient required intra-operative surgical conversion and subsequently died. – one patient required intra-operative surgical conversion and experienced temporary mesenteric and renal ischemia. – one patient required an additional surgical intervention and experienced temporary renal ischemia. – one patient sustained no injuries due to deployment during an open repair. • while incomplete deployments are known adverse events and identified within the instructions for use (ifu), gore has seen an increased frequency of these partial deployment events in conformable tag® devices sold (totaling 0.03% of 12,865 devices distributed*) that were manufactured in the prior year compared to those manufactured earlier. .

Manufacturer