Notificaciones De Seguridad De Campo acerca de DRX-Revolution

Según Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), este evento ( notificaciones de seguridad de campo ) involucró a un dispositivo médico en Slovenia que fue producido por Carestream Health.

¿Qué es esto?

Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • Fecha
    2017-01-13
  • País del evento
  • Fuente del evento
    AMPMDRS
  • Notas / Alertas
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notas adicionales en la data
  • Causa
    Carestream health has identified a defect in the production of the drx-revolution that could cause it to move inadvertently. in our own investigation of the only reported event, we found that the center lever of the drive lever, which had escaped and had an effect on the operation of the drive and brake mechanism, prevented the device from stopping. there were no consequential injuries, but there is a risk that the device may have a life-cycle problem with other units. failure to stop the unit and contact with the person may result in injury.

Device

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    AMPMDRS