Notificaciones De Seguridad De Campo acerca de Elecsys 2010 analyzer

Según Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), este evento ( notificaciones de seguridad de campo ) involucró a un dispositivo médico en Slovenia que fue producido por Hitachi High Technologies Corporation (HHT).

¿Qué es esto?

Las notificaciones de seguridad de campo son comunicaciones sobre acciones que se pueden estar tomando en relación con un producto que se encuentra en el mercado, enviadas por los fabricantes de dispositivos médicos o sus representantes. Estas notificaciones son principalmente para trabajadores de la salud, pero también para usuarios. Pueden incluir alertas y retiro de equipos.

Más información acerca de la data acá
  • Tipo de evento
    Field Safety Notice
  • País del evento
  • Fuente del evento
    AMPMDRS
  • Notas / Alertas
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • Notas adicionales en la data
  • Causa
    We regret to inform you that in very rare cases a software (sw) malfunction in the sample&control datafile can occur which may lead to a potential data mismatch. this sw malfunction only occurs: ∑ when the “sample data clear” function is not performed daily as indicated in the operator’s manual, and ∑ when the sample&control datafile of the elecsys 2010 analyzer is filled with > 600 records. roche has received no customer complaints for this issue. this error only occurs if deletion of documented results or samples is not performed as pre or post operation function. .

Manufacturer

  • Empresa matriz del fabricante (2017)
  • Source
    AMPMDRS