Retiro De Equipo (Recall) de Baxter Lock Box for 250 mL Container

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Baxter Healthcare Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    29958
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1570-05
  • Fecha de inicio del evento
    2004-09-01
  • Fecha de publicación del evento
    2005-04-22
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2006-07-21
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Pump, Infusion - Product Code FRN
  • Causa
    The pump may free-flow if the door to the 6060 pump is not closed prior to loading the pump in to the lock box. the pump will alarm 'door open', but if the pump is turned off to silence the alarm while remaining in the lock box, the fluid will free-flow.
  • Acción
    Urgent Device Correction letters dated 9/1/04 were sent to all U.S. Baxter affected customers on the same date, to the attention of the Director of Nursing. The facilities were advised of the free-flow incident and as a result that Baxter wanted to heighten awareness of the proper clinical procedures for loading the sets into the infusion systems using these Lock Boxes with the 6060 Multi-Therapy Pump. Baxter''s investigation into the event confirmed that the 6060 pump door was left open after the lock box was attached to the pump. Baxter provided copie of the instructions for use of the lock box and pages from the 6060 pump operator''s manual which outline the proper procedure for loading the sets into the the pump and the pump into the lock box. The accounts were requested to complete and fax back to Baxter the enclosed response form, erifying receipt of the letter, confirming that the information has been provided to all users of the infusion system, and allowing Baxter to order and ship labels for the lock box and revised instructions for use to the facility. Any questions were directed to 1-800-843-78867.

Device

  • Modelo / Serial
    all lot numbers/serial numbers of product codes 2L9363, 606160 and 2M9832P
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide and internationally to Australia, Belgium, Canada, France, Hong Kong, Portugal, Spain and the United Kingdom.
  • Descripción del producto
    Baxter Lock Box for 250 mL Container, for use with 6060 Pump Family; Baxter product code 2L9363 and previously marketed by Sabratek under product code 606160, Baxter Healthcare Corporation, Deerfield, IL 60015; Product code 2L9363 is sold separately and as a component of the 6060 Pain kit, product code 2M9832P.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Baxter Healthcare Corp., Rt. 120 & Wilson Rd, Round Lake IL 60073
  • Source
    USFDA