Retiro De Equipo (Recall) de Device Recall Kelyniam CSI

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Kelyniam Global, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    79791
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1872-2018
  • Fecha de inicio del evento
    2017-06-10
  • Estado del evento
    Open, Classified
  • País del evento
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Plate, cranioplasty, preformed, non-alterable - Product Code GXN
  • Causa
    Significant design changes made to the custom skull implant (csi) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to fda for marketing clearance. the temporal suture system (tss) is no longer being offered and the integrated fixation system (ifs) has been temporarily suspended.
  • Acción
    On June 10, 2017, the recalling firm notified its Sales Representatives that effective June 10, 2017, the firm would no longer offer the temporal suture system (TSS). The firm requested that sales reps dispose of all marketing materials which contain details regarding the TSS. Sales reps who wanted to keep existing booklets in the field were asked to contact the firm to obtain adhesive labels to place over the discontinued product. On November 13, 2017, the recalling firm notified its Sales Representatives that the firm was temporarily suspending shipment of the Integrated Fixation System (IFS). The firm stated that all paper brochures were to be discarded, and asked sales reps to inform all those who received "Customize Skull and Craniofacial Implants" brochure and IFU. Revised marketing materials are available on the firm's website, www.kelyniam.com. If you have any questions, you may contact Dr. Mark V. Smith, Vice President of Kelyniam Global, Inc., at msmith@kelyniam.com, or by phone at (800) 280-8192 x702.

Device

  • Modelo / Serial
    Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C.  Case Numbers: CSI111715-KH1 CSI072814-JB1 CSI082514-MA1 CSI080714-CT1 CSI091914-JW1 CSI101614-TR1 CSI041415-MZ1 CSI100915-DD1 CSI120215-MM1 CSI100915-DD2 CSI011816-TA1 CSI012016-RT1A CSI012716-MP1 CSI021116-CA1 CSI030716-BC1 CSI040716-CK1 CSI060916-DF1 CSI070716-AC1 CSI092116-JA1 CSI110416-DW1 CSI120616-LS1 CSI012417-SD1 CSI062617-HR1
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    US nationwide distribution.
  • Descripción del producto
    Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Kelyniam Global, Inc., 97 River Rd, Collinsville CT 06019-3246
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA