Retiro De Equipo (Recall) de Device Recall Merge RadSuite software

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Merge Healthcare, Inc..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    76258
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1180-2017
  • Fecha de inicio del evento
    2015-11-17
  • Fecha de publicación del evento
    2017-02-09
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2017-02-10
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, image processing, radiological - Product Code LLZ
  • Causa
    The values provided from the pixel value tool do not appear to be correct, which may result in potential patient injury or delay in diagnosis or treatment.
  • Acción
    Merge sent an Urgent Medical Device Recall letter dated March 28, 2016 to all affected customers. The letter notifies the customer of the issue and informs them an upgrade is available. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they are interested in accepting the fix. If they decline the fix, they are asked to provide the reason why. The response was required no later than April 18, 2016. A second letter dated July 11, 2016, was issued asking for a response by August 2, 2016, even though all upgrades were made because the effectiveness checks were not yet completed. The letter and response form contained the same information as the first letter. The recalling firm reported emails were tracked for delivery confirmation and the letters were tracked for delivery by USPS certified services. For further questions, please call 877-744-5369.

Device

  • Modelo / Serial
    Version V8.30.7.8
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    US Distribution was made to medical facilities in AL, IN, MI, PA and WI.
  • Descripción del producto
    Merge RadSuite software. The firm name on the label is Merge Healthcare, Hartland, WI.
  • Manufacturer