Retiro De Equipo (Recall) de Device Recall NeuViz Dual series computed Tomography Scanner System

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Philips And Neusoft Medical Systems Co., Ltd..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    48727
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2307-2008
  • Fecha de inicio del evento
    2008-06-23
  • Fecha de publicación del evento
    2008-09-17
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2010-09-24
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, x-ray, tomography, computed - Product Code JAK
  • Causa
    Potential for r-host box and components inside to be detached of its mounting during the gantry rotor rotation. there is a potential for components to be damaged inside the system gantry and for the internal components to be expelled from the system gantry and injure the patient, operator, bystander or service people.
  • Acción
    On 6/23/08/08, the Philips Neusoft sent a letter dated 6/10/08 to their US customers explaining the problem, under what circumstances it can occur, and the actions that should be taken by the customer/user in order to prevent risks for patients or users. The field correction is install a R-host bracket to aid supporting of the R-host computer. The letter included additional information to customers who had already received the field action (this was an on-going activity that was re-classified as a potential safety). Customers did not receive the additional information about the status of their upgrade if the upgrade had not occurred at the time of the mailing. They were informed that they would be contacted by Newsoft Medical systems representative for the planning of this repair. If you have questions contact Service Support Department at nms-service@neusoft.com.

Device

  • Modelo / Serial
    Serial numbers 400562, 400563, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH025EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, and NDH034EI
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    The units were distributed to one distributor in NC, and further distributed to medical facilities located in CO, FL, KY, NC, SC, TN and TX.
  • Descripción del producto
    NeuViz Dual series computed Tomography Scanner System, X-Ray System