Retiro De Equipo (Recall) de Device Recall PerformanceLOAD Ambulance Cot Fastener

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Stryker Medical Division of Stryker Corporation.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    75820
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0739-2017
  • Fecha de inicio del evento
    2016-11-08
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2018-01-16
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Stretcher, wheeled - Product Code FPO
  • Causa
    Stryker medical initiated a voluntary recall of stryker performance-load cot fasteners due to complaints that fastening system may not have been securely fastened or had an inability to fasten into the performance-load and bounced back during loading which could cause injury to consumers.
  • Acción
    Stryker Medical initiated a voluntary recall of Stryker Performance-LOAD Cot Fasteners MODEL: 6392, due to complaints and service reports that some cots compatible with the Performance-LOAD cot fastening system may not have been securely fastened or had an inability to fasten into the Performance-LOAD (bounced back during loading) which could lead to injury of consumers. Customer notification letters were sent via Fedex on November 8, 2016. Customers were asked to quarantine the impacted devices and submit completed business reply forms by fax (269)488-8691 or email productfieldaction@stryker.com. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the modification/repair of your Performance-LOAD unit to ensure that the unit securely fastens the cot within the ambulance. 4. In the interim, until the service can be performed, there are additional actions that users can and should take to ensure that the cot is securely fastened into the Performance-LOAD system. a. Per the PerformanceLOAD Operations Manual (P/N 6392-009-001, pages 26-27. If customers have loaned or sold any of the products listed in this letter, please forward a copy of the recall notice to the new users and advise Stryker of their new location in the space provided on the business reply form. If you have any questions or concerns, please contact Stryker Customer Service (1-800-327-0770). Our normal business hours are Monday-Friday 8 a.m-6 p.m. (EST).

Device

  • Modelo / Serial
    Model(s): 6392-000-001, 6392-000-002, 6393-000-003, 6393-000-004; GTIN: 07613327282542, 07613327282603, 07613327282597, 07613327282610
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution to AK, AL , AR , AZ , CA , CT , DC , DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA , RI, SC, TN, TX, UT,VA, VT ,WA ,WV ,WY and Internationally to Canada and the Netherlands
  • Descripción del producto
    Performance-LOAD Ambulance Cot Fastener || Stretcher, Wheeled
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Stryker Medical Division of Stryker Corporation, 3800 E Centre Ave, Portage MI 49002-5826