Retiro De Equipo (Recall) de Device Recall S TBone Plate; M TCompression Plate;

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Stryker Howmedica Osteonics Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    50094
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1334-2011
  • Fecha de inicio del evento
    2006-05-26
  • Fecha de publicación del evento
    2011-02-16
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2014-12-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Plate, fixation bone - Product Code HRS
  • Causa
    A package labeled as a s t-bone plate, catalog number 57-05260, lot code g4m00f1576, may actually contain a m t-compression plate, catalog number 57-10250, lot code g4n00f1584 and vice versa.
  • Acción
    Stryker Orthopedics sent an Important Market Withdrawal letter dated May 26, 2006, by Federal Express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory, including their hospital consignment location, to identify the product. If the affected product remained in its original packaging, customers were instructed to retrieve all of the affected product lots and return it to their branch or agency warehouse for reconcillation. Reconcile all product on the attached Product Accountability Form and fax a copy of the form to 201-831-6069 within 5 days of receipt of the notice. Customers were asked to return the affected product via Inter-Org Transfer and ship to Product Surveillance, Stryker Orthopaedics, 325 Corporate Drive, Mahwah, New Jersey, USA 07430-2104. Customers were also instructed to contact their Stryker Orthopaedics Customer Service representataive to re-order the product that they were returning. For questions regarding this recall call (201) 831-5825.

Device

  • Modelo / Serial
    S T-Bone Plate, Product number: 57-05260; Lot Code G4M 00F1576; M T-Compression Plate, Product number 57-10260; Lot Code G4N00F1584
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    Nationwide Distribution
  • Descripción del producto
    S T-Bone Plate; || M T-Compression Plate; || Manufactured and Distributed by Stryker Lebinger GmbH & Co. || Freiburg, Germany. || The Profyle System is intended for use in internal fixation of the bones of hand and wrist. Examples of these procedures may include but are not limited to replantation, lag screws techniques, joint fusion, corrective osteotomies, and the treatment of fractures
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Stryker Howmedica Osteonics Corp., 325 Corporate Dr, Mahwah NJ 07430
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA