Retiro De Equipo (Recall) de Device Recall SSH880CV Aplio Artida

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Toshiba American Medical Systems Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    55543
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2203-2011
  • Fecha de inicio del evento
    2010-01-06
  • Fecha de publicación del evento
    2011-05-19
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2012-04-20
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, imaging, pulsed doppler, ultrasonic - Product Code IYN
  • Causa
    Toshiba america medical systems (tams) has become aware that the flex m software has reversed left and right annotation and post flex m measurement error, which may result in incorrect values.
  • Acción
    The recall communication was initiated on 1/6/10 with Toshiba forwarding an Urgent: Medical Device Correction letter with attached Customer Notification Reply Form (via US Postal Service for US Customers) to all customers who purchased the FLEX-M mode on Aplio Artida Diagnostic Ultrasound System. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to avoid performing M image measurement using images saved in FLEX-M mode until the corrective action is completed. If customers need to perform measurement for images saved in FLEX-M mode, they were instructed to use manual calibration. Customers were instructed to note that even though the corrective action is taken, the existing saved images (saved images before the corrective action is taken) are not modified. Therefore, when referring to the measured values on the images which are saved in FLEX-M mode before the corrective action is taken, customers were instructed to carefully handle the measurements results because they may be incorrect. Customers were instructed to complete the attached form and fax it to the toll free number at the top of the form. The form can also be sent via e-mail to raffairs@tams.com. If customers have any questions regarding the medical device correction letter, they can contact the Director of Regulatory Affairs at (800) 421-1968 or their local Toshiba representative.

Device

  • Modelo / Serial
    W1A0822031-33, W1A0822048-49, W1A0832056, W1A0842069-70, W1B0852071-80, W1B0862087-8, W1C0872103-08, W1C0882119-24, W1F0972204-06, W1F0972218-20, W1F0992260-62, W1F0992268, and W1F0992270-73.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Nationwide Distribution -- Including states of IA, MT, PA, CA, OR, GA, LA, OH, TX, MA, OH, WI, and Puerto Rico.
  • Descripción del producto
    Software for SSH-880CV Aplio Artida; System Diagnostic Ultrasound
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Toshiba American Medical Systems Inc, 2441 Michelle Dr, P.O. Box 2068, Tustin CA 92781-2068
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA