Retiro De Equipo (Recall) de Device Recall Stryker Electric Stretcher, Model 1050.

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Stryker Medical Div. of Stryker Corporation.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    51223
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1230-2009
  • Fecha de inicio del evento
    2009-02-27
  • Fecha de publicación del evento
    2009-04-05
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2011-04-25
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Wheeled Hospital Stretcher - Product Code FPO
  • Causa
    The head section backrest (fowler) may not raise up, or, if up, may drop suddenly due to wearing out of the actuator threads on the fowler.
  • Acción
    Consignees were notified by an Urgent - Medical Device Correction letter dated 2/27/09, advised of the problem and instructed to locate stretchers and verify that there is no grinding noise when activating the fowler electrically. Consignees were also instructed to remove any stretchers with a fowler that was non-operational, or which makes a grinding sound, from service until the firm makes repairs and contact the firm. Stretchers will be corrected within the next four months by a Stryker field service representative. The enclosed post card should be returned to confirm receipt of the notification. The notice should be forwarded to any users that have borrowed or purchased the stretchers and the firm should be notified of their new location. Also, if any of the affected stretchers have been disposed of or are no longer in use, the firm needs to be provided with their serial numbers. Questions or concerns should be directed to Andrea Kuthe at 269-324-6581.

Device

  • Modelo / Serial
    Serial numbers 94113214, 94113215, 94113216, 94113217, 94113218, 94113219, 94113220, 94113221, 94113222, 94113223, 94113224, 94113225, 94113226, 94113227, 94113228, 94113229, 94113230, 94113231, 94113232, 94113233, 94113234, 94083551, 94083552, 94083553, 94083554, 94083555, 94083556, 94083557 A, 94083558, 94113211, 9501031208, 9501031209, 9501031210, 9501031211, 9503032237, 9503032238, 9503032239, and 9503032240.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Worldwide Distribution
  • Descripción del producto
    Stryker Electric Stretcher with the Quick Drop CPR option, Model 1050.