Retiro De Equipo (Recall) de Device Recall Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Stryker Howmedica Osteonics Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    74619
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2390-2016
  • Fecha de inicio del evento
    2016-06-23
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2018-08-01
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/ - Product Code MBH
  • Causa
    It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the specialty triathlon tibial alignment handle with secondary lock assembly during surgery. upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.
  • Acción
    Stryker sent an Urgent Product Recall letter dated June 24, 2016, with a Business Reply Form attached to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to inform users of the Urgent Product Recall letter and forward the notice to all those individuals who need to be aware within their organization. Customers were instructed to return all affected product at their location to: Stryker C/O Stericycle 2670 Executive Dr., Suite A Indianapolis, IN 46241 Customers were also instructed to complete and sign the enclosed Business Reply Form and fax it to 888-912-7352 or email to Stericycle at strykerortho4582@stericycle.com. Customers with questions were advised to call (201) 831-6693. For questions regarding this recall call 201-831-5000.

Device

  • Modelo / Serial
    Catalog No: I-K3254TA00 Lot Nos: F6E13243, F6L13864, F6M14078, F6S14667, F6S14831, F6W15444, F6W15406, F7C15943, F7H16099 and F7L16994
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT
  • Descripción del producto
    Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, || Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Stryker Howmedica Osteonics Corp., 325 Corporate Dr, Mahwah NJ 07430-2006