Retiro De Equipo (Recall) de Device Recall Stryker Vitallium Wire (6 Pack) Non Sterile

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Stryker Howmedica Osteonics Corp..

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    73042
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0855-2016
  • Fecha de inicio del evento
    2015-12-23
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2017-05-19
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Staple, fixation, bone - Product Code JDR
  • Causa
    The wire packages are correctly marked with a "non-sterile" label, however, the enclosed instructions for use (ifu) states the device are sterilized via gamma irradiation and should not be resterilized. since the ifu states the devices are sterile, no instructions for moist heat sterilization are provided.
  • Acción
    Stryker Branches/Agents were notified of the action via email on December 23, 2015. Urgent Medical Device Recall Notification Letters/Urgent Medical Device Recall Notification Acknowledgement Forms dated December 23, 2015 were sent to Branches/Agents via UPS on December 28, 2015. Urgent Medical Device Recall Notification Letters/Urgent Medical Device Recall Notification Acknowledgement Forms dated December 23, 2015 were sent to Hospital Risk Managers and doctors via UPS on 12/23/2015. The notification instructed customers of the related issue with the affected product; how to identify affected product; the potential hazards associated with affected product; risk mitigations; and actions needed to be taken. Customers were asked to complete and return the attached Product Recall Acknowledgement Form within 5 days and either email (strykerOrtho7808@stericycle.com or Fax (1-866-672-0627 ) the response form back. Customers were instructed to return all affected products to Stryker C/O Stericycle, 2670 Executive Drive, Suite A, Indianapolis, IN 46241 (Attn RA2015-120-Event 7808). A point of contact was provided in case the customer had any questions 201.831.5272 .

Device

  • Modelo / Serial
    All lots of the following Catalog Numbers: 2704-1-024; 2704-3-018; 6703-1-018; 6703-2-120; 6704-1-018; 6704-2-120 and 6704-3-120.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    Yes
  • Distribución
    Nationwide Distribution.
  • Descripción del producto
    Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures.
  • Manufacturer

Manufacturer

  • Dirección del fabricante
    Stryker Howmedica Osteonics Corp., 325 Corporate Dr, Mahwah NJ 07430-2006
  • Empresa matriz del fabricante (2017)
  • Source
    USFDA