Retiro De Equipo (Recall) de The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.

Según U.S. Food and Drug Administration, este evento ( retiro de equipo (recall) ) involucró a un dispositivo médico en United States que fue producido por Arrow International Inc.

¿Qué es esto?

Una corrección al equipo o acción de retiro tomada por el fabricante para abordar un problema con un dispositivo médico. Los retiros (recalls) ocurren cuando un dispositivo médico está defectuoso, cuando puede poner en riesgo la salud, o cuando simultáneamente está defectuoso y puede poner en riesgo la salud.

Más información acerca de la data acá
  • Tipo de evento
    Recall
  • ID del evento
    31765
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0842-05
  • Fecha de inicio del evento
    2005-03-18
  • Fecha de publicación del evento
    2005-05-13
  • Estado del evento
    Terminated
  • País del evento
  • Fecha de finalización del evento
    2006-11-22
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter, Embolectomy - Product Code DXE
  • Causa
    The orange pebax lumen located within the catheter basket may separate from the catheter basket sleeve permitting the lumen to become dislodged during use.
  • Acción
    The firm began notification of consignees on 3/23/2005 via certified letter.

Device

  • Modelo / Serial
    All units with Lot Numbers beginning with: MF311; MF312; MF401; MF402; MF403; MF404; MF405; MF406; MF407; MF408; MF409; MF410; MF411; MF412; MF501.
  • Clasificación del producto
  • Clase de dispositivo
    2
  • ¿Implante?
    No
  • Distribución
    Product was distributed nationwide to dealers, hospitals, sales representatives, and subsidiaries (distributors). The product was also distributed to Australia, Canada, Colombia, Czech Republic, France, Germany, Israel, Italy, Japan, Netherlands, Russia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey, United Kingdom. There are no government accounts.
  • Descripción del producto
    The Arrow Trerotola Percutaneous Thrombolytic Device (Product Codes: PT-65709-WC, PT-65709-W, PT-65709-HFWC, and PT-12709-WC) consists of a catheter with an expandable fragmentation basket attached to a through lumen torque cable and an outer catheter sheath. The catheter is used in conjunction with a low-speed, battery operated rotator unit to treat thrombosed hemodialysis grafts and AV fistulas. The macerating mechanism is done by 9 mm basket (made up of four stranded stainless steel wires) that rotates at 3000 rpm when rotator is activated. The current device is 7 Fr in size and comes in two lengths -- 65 cm and 120 cm. This device is used by inserting the catheter into the occluded graft or fistula via introducer sheath and then activating the rotator unit.