SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The problem associated with the affected product is the risk of stress cracks developing in the blade
tip that may not be readily visible during routine inspection prior to intubation. this cracking may
eventually cause the product to fail.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Vascutek has received a number of events involving graft bleeding including anastomotic bleeding and suture hole bleeding, either during or shortly after surgery.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
This notice is to inform you about important updates to the above products and their instructions for use (ifu).
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
When an image sequence template is added to a field in rt chart and a ciao (completely
irradiated aperture opening) is calculated, the
resulting ciao display at 4ditc may be incorrect.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
May exhibit unexpected 6mv beam output variations.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Varian medical systems has identified a problem in aria oncology information system (ois) for radiation oncology such that a user may inadvertently chart medical record information in an unintended patient’s chart.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
A software anomaly has been identified in truebeam when using respiratory gating with patients treated in the prone position. the anomaly is the respiratory gating curve is inverted when a prone patient plan is loaded into truebeam.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The rectal retractors have shown evidence that contaminates, or fluids, have entered into the center of the rectal retractor paddle through the cross pins or bottom rod areas. using a part in this condition could result in infection, toxicity, or allergic reaction.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR, MHRA
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
This advisory notice is to provide you with updated information on multi-isocenter plan management and acquisition of couch parameters for multi-isocenter plans at the 4ditc.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Varian states that excessive connector resistance on the above devices may cause the jaw positions to differ from the intended jaw positions without warning the operator, potentially affecting treatment.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
The 4d integrated treatment console (4ditc) is scheduled for
a mandatory upgrade from your current version to 10.2. the reason for the upgrade is to
ensure customers have the most current version of 4ditc which contains several
improvements to address issues related to proper recording of treatment information from
varian machines.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
While operating with the 6-dot marker block, above version of rpm gating may produce an inverted respiratory waveform in amplitude and phase-gating sessions.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
The wedge angle labeling on the wedge tray may not correctly match the wedge body. in all cases the
wedge body is labeled with the correct wedge angle.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
A possibility of a varian mlc interfering with the treatment field of a demountable brainlab m3 micro-mlc without an interlock being asserted.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
An outdated assigned hounsfield unit [hu] value can be used for dose calculation when the clock on the eclipse client workstation is not synchronized with the database/system server.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
The treatment field energy may change from the current selection to
a previous selection without user intervention and without invalidating the mu or calculated dose
distribution.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
In affected eclipse versions, user editing of mlc leaf positions may not automatically invalidate calculated dose distributions.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Varian has identified an installation anomaly whereby the bolts used to fasten the counterweight to the gantry may not, in some cases, be fully tightened to the required torque specification. in the worst possible scenario this could result in the counterweight detaching from the gantry; and the resulting unbalanced gantry may swing freely, posing a hazard of a collision with a patient, therapist, or service person.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
A system was discovered where vmat (volumetric modulated arc therapy) plans were delivered as dynamic arc. it was determined that an older configuration of the clinac and older version and configuration of the linac vi software which did not recognize vmat plans was inadvertently installed.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Admin instructions may not print on drug order prescriptions or be transmitted with e-rx. when the user changes the prescription “type” field after inputting admin instructions in the text box, the admin instructions portion of the prescription will not be saved to the database. an e-rx transmitted to the pharmacy will not contain the intended admin instructions.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Varian medical systems has received reports of an anomaly that can result in a treatment without intended gating. this issue occurs when a patient planned with gating is treated on more than one truebeam system [includes: truebeam, truebeam stx, vitalbeam or edge]. varian has not received any reports of injury due to this issue. the issue only affects truebeam systems version 2.0, 2.5 and 2.5 mr1.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
If the dose prescription of a vmat or rapidarc plan is changed after the initial dose volume optimization, the monitor units and dose distribution may not reflect the intended prescription. affects eclipse v8.5 to v11.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The shaft of the type 01 elbow motor for exact arms used in the on-board imager® [obi] device have failed due to belt over-tensioning. this over tensioning would have taken place during servicing. this failure would allow the exact arm to swing freely with gravity.
no injuries have been reported to-date. only type 01 elbow motors are subject to these potential failures, type 02 elbow motors are not affected.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The plan parameters workspace within eclipse™ treatment planning system, aria® radiation oncology, aria® radiation therapy management, and acuity™. when a dose relevant parameter is changed and “do not clear mus and reference point dose” is selected, the monitor units (mus) may change.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
Anomaly in eclipse versions 8.0 to 8.6 whereby a
particular plan approval workflow may result in a difference between the intended ssd value
and the ssd value communicated to the treatment delivery system.