Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Working channel sleeve may protrude from camera cap, potentially damaging duct walls.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Arcomed ag has been informed that in some rare cases the syramed μsp6000 and volumed μvp7000 would signal a safety alarm. during the alarm the devices stop the infusion until restarted by the user. some reports indicate that on some pumps the display of the time and date was not correct. this is only an informative display and has no influence on the function of the pump.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Arjo has received a low number of complaints claiming there were difficulties with releasing the buckle to gain access to the patient.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
An incorrect routing of a mains supply cable may allow for the supply cable to be cut by the high/low actuator yoke. a cut mains supply cable may cause an electrified framework leading to the risk of serious injury or death.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company
Causa
The manufacturer conducted field safety correction actions in canada between 2004 and 2007 to provide improved safety instructions and warnings to reduce the risk of the alenti chair tipping and patient falls. the incident rate has since declined. however, the manufacturer continues to receive several incident reports in canada each year. since august 2012, there have been nine tipping and patient slide-out incidents reported in canada. two of these were associated with patient deaths ( more details in attached file ).
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Arjo received several customer complaints on powerdrive system in citadel plus beds claiming following issues: the device moves forward at maximum speed with only left trigger squeezed, drives without handles being held, moves forward even when the right hand trigger is released.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The lock pin can become stuck in the "in" position during the manual rotation, resulting in the inability to prone a patient.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Arjohuntleigh states that the seat/stretcher of the above pool lifts may detach if the user fails to dock the chair or stretcher correctly, potentially resulting in patient falls.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Arrow international is recalling the product referenced above because the packaging may not be sealed. if the packaging is compromised in this manner, the sterility of the product cannot be guaranteed.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Arrow is recalling certain lots of the products referenced above due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
A sharp edge on the cannulated tip of the device may damage or cut a suture.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
When assignments are changed for a location, the location conditions are stored in amemory cache; however, these entries are not deleted. eventually the memory cache becomes full which results in a module reboot. during this time, no information is communicated to any display devices.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Please be advised that aspen medical has become aware of an issue related to the application of an incorrect expiry date to a number of sorbsan dressings.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Medical noted that the sealing of the outer blister can be weakened.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCAR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Two boxes of surgical sutures were stolen in colombia before sterilization. the products are marked ce and sterile, however, the stolen material was not sterilized yet.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
This recall has been initiated in response to a seal defect found in certain sterile barrier pouches containing the device. affected pouches are un-sealed at the bottom.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
This letter is intended to provide clarification about the current packaging configuration of the atrium ocean, oasis, express and express mini chest drainage products. atrium chest drainage products are packaged with a double sequential sterile wrap and placed into an outer dust cover. the dust cover is folded over and secured with the package label.
this packaging system is the same configuration used on atrium chest drainage products for the past 30 years. this field notification is to provide clarification as to which part of the packaged drain is sterile.
the label on the outer package denotes the product as being sterile. the chest drain inside the packaging is sterile; the outer dust cover is not sterile. the two sterilization wraps are what provide the sterile barrier for the product.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The auxiliary o2 & suction module sometimes can’t be activated by the on/off switch. the auxiliary o2 part of the module is not affected. there are no reported injuries.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
In b braun manufacturing site it was determined that the irradiation dose qualified for sterilization of the above mentioned products was too low. in consequence, the germ reduction through gamma irradiation may not have reached the requested sterility assurance level of 10-6. the effect cannot be limited to specific batches.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
In certofix quattro minor inter-lumen connections may occur with a very low frequency.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: BfArM
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
During continuous post-market surveillance we became aware of a disconnection of the female luer-lock connector from the venous side of the single-needle adapter leading to blood loss for the patient.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The perifix catheter connector is a connection device used by clinicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. the connector is used in conjunction with the catheter for continuous administration of anesthetic fluids. in the course of our regular post market surveillance activities we have found that the perifix catheter connector may not remain closed during use. in some cases this has led to leakage or disconnection of the catheter from the perifix catheter connector.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
Leakages at certain bicarbonate and end conductivity sensors might result in uf balance deviations.
Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
Notas adicionales en la data
Report Source: NCMDR
SFDA urges all hospitals that have devices subjected to recall, to contact the company.
Causa
The ingredient that gives hydrophilic properties to this impression material was left out of the formulation. the two different trade names and corresponding lot numbers are listed above.