Retiro De Equipo (Recall) de Device Recall VITROS 4600 Chemistry System
  • Tipo de evento
    Recall
  • ID del evento
    69830
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0968-2015
  • Fecha de inicio del evento
    2014-11-19
  • Fecha de publicación del evento
    2015-01-14
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2018-05-21
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131870
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Causa
    Software anomaly: the firm has identified an anomaly with vitros system software version 3.1 and below, and determined that the software may not properly identify an expired calibration.
  • Acción
    On 11/19/2014, IMPORTANT PRODUCT CORRECTION NOTIFICATION letter (Ref. CL2014-296, dated 11/19/2014) was sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to the US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and advise them of the required actions. Immediately install the most recent ADD using the All Assay Data option. If any of your calibrations are expired, the system will post PZB-005 condition codes. Perform calibrations prior to using those reagents. As expected, calibrations that will expire within one hour will also be flagged. Continue to install the most recent ADD at least once each week using the All Assay Data load option so that the system properly flags expired calibrations. --- Foreign affiliates were informed by e-mail on 11/19/2014 of the issue and instructed to notify their consignees of the issue and required actions.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, NEEDLE NOSE FORCEPS, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0780-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131871
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, SCHERTL GRASPER, FENESTRATED, 29CM,
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0781-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131872
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, BLUNT WITH SPOON, GRASPER, 29CM,
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0782-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131873
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall VITROS 5600 Integrated System
  • Tipo de evento
    Recall
  • ID del evento
    69830
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0969-2015
  • Fecha de inicio del evento
    2014-11-19
  • Fecha de publicación del evento
    2015-01-14
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2018-05-21
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131874
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Causa
    Software anomaly: the firm has identified an anomaly with vitros system software version 3.1 and below, and determined that the software may not properly identify an expired calibration.
  • Acción
    On 11/19/2014, IMPORTANT PRODUCT CORRECTION NOTIFICATION letter (Ref. CL2014-296, dated 11/19/2014) was sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to the US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and advise them of the required actions. Immediately install the most recent ADD using the All Assay Data option. If any of your calibrations are expired, the system will post PZB-005 condition codes. Perform calibrations prior to using those reagents. As expected, calibrations that will expire within one hour will also be flagged. Continue to install the most recent ADD at least once each week using the All Assay Data load option so that the system properly flags expired calibrations. --- Foreign affiliates were informed by e-mail on 11/19/2014 of the issue and instructed to notify their consignees of the issue and required actions.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, BLUNT GRASPER, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0783-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131875
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, BULLET FORCEPS, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0784-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131876
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, MARYLAND, FORCEPS CURVED, 29CM,
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0785-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131877
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, BIOPSY FORCEPS, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0786-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131878
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, BIOPSY PUNCH, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0787-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131879
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, STRAIGHT NEEDLEHOLDER, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0788-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131880
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, DORSEY GRASPER, 20CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0789-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131881
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, DORSEY GRASPER, 29CM,
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0790-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131882
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, ATRAUMATIC GRASPER, SINGLE ACTION, 29CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0791-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131883
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, ATRAUMATIC GRASPER, DOUBLE ACTION, 20CM
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0792-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131884
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    The parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM INSERT, ATRAUMATIC GRASPER, DOUBLE ACTION, 2...
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0793-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131885
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    Stryker endoscopy is recalling laparoscopic manual instruments and accessories because the parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG., 3MM INSERT, BABCOCK FORCEPS, 29CM,
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0794-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131886
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    Stryker endoscopy is recalling laparoscopic manual instruments and accessories because the parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM X 20CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW,
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0795-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131887
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    Stryker endoscopy is recalling laparoscopic manual instruments and accessories because the parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM X 29CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0796-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131888
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    Stryker endoscopy is recalling laparoscopic manual instruments and accessories because the parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, 3MM X 29CM INSERT, DISSECTING FORCEPS
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0797-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131889
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    Stryker endoscopy is recalling laparoscopic manual instruments and accessories because the parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall PKG, CANNULA 8MM AUTOMATIC VALVE STOP COCK, W/O TROCAR
  • Tipo de evento
    Recall
  • ID del evento
    69838
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0798-2015
  • Fecha de inicio del evento
    2014-11-17
  • Fecha de publicación del evento
    2014-12-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131890
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Endoscope and/or accessories - Product Code KOG
  • Causa
    Stryker endoscopy is recalling laparoscopic manual instruments and accessories because the parameters provided in the laparoscopic manual instruments instructions for use (ifu 1000-401-070 revision g or prior) do not support the unwrapped gravity cycle and the ethylene oxide cycle sterilization methods.
  • Acción
    Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Retiro De Equipo (Recall) de Device Recall ARTISTE MV System, LInear Accelerator (LINAC)
  • Tipo de evento
    Recall
  • ID del evento
    69880
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0984-2015
  • Fecha de inicio del evento
    2014-11-19
  • Fecha de publicación del evento
    2015-03-18
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-04-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131894
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Accelerator, linear, medical - Product Code IYE
  • Causa
    There may be an existing dark current phenomenon on artiste linac in combination with imrt or marc treatments using unflat beams. software issue.
  • Acción
    Siemens Planned Action: 1. Providing recommendations for user actions in terms of routine dosimetric quality control checks for any IMRT treatment at very high dose rates (Multiple X) as well as for mARC treatments. 2. Software update for the Function Controller #3, which controls parts of the beam generation. This software update enables the correct switching between Pulse Forming Network (PFN) and reduced Pulse Forming Network (iPFN) in pause state, to reduce the dark current during the movement as much as possible. This software will be installed by a Siemens service representative. 3. Siemens will contact the user to make an appointment to implement the modification system at a date and time when the system will be available. These corrections shall be made free of charge. The deployment of the permanent correction shall be completed by July 1, 2015. For further questions please call (610) 219-6300.
Retiro De Equipo (Recall) de Device Recall Phadia 1000 Instrument
  • Tipo de evento
    Recall
  • ID del evento
    69872
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0947-2015
  • Fecha de inicio del evento
    2014-11-20
  • Fecha de publicación del evento
    2015-01-12
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-06-30
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131895
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, test, radioallergosorbent (rast) immunological - Product Code DHB
  • Causa
    During an investigation of instrument logs it was determined that in specific circumstances involving multiple steps, a rack sequencing error may occur. this will result in a mismatch between the sample id and the test result reported for all subsequent sample racks in that run.
  • Acción
    ThermoFisher sent an Urgent Medical Device Product Correction letter to all affected customers. The firm will inform customers of the specific sequence of errors and operator actions that can trigger the event via letter and phone, and will request that they do not respond to the Barcode Read Error and let the sample with the error eject normally from the instrument and be rerun after the barcode read error has been resolved. The firm will issue a mandatory Phadia 1000 Instrument Software (ISW) update that will correct the issue and make efforts to install it within 45 days from date of availability. Customers were asked to return the written response form via fax to: 1-888-243-5214 Attention: Customer Support, Phadia 1000 ISW Product Correction Response; EMAIL: Compliance-us.idd@thermofisher.com Email Subject Line: Phadia 1000 ISW Product Correction Response. Customers with questions were instructed to contact Phadia US Technical Support at 800-346-4364, option 2. For questions regarding this recall call269-492-1957.
Retiro De Equipo (Recall) de Device Recall Biograph mCT Flow Edge4R
  • Tipo de evento
    Recall
  • ID del evento
    69897
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0807-2015
  • Fecha de inicio del evento
    2014-11-13
  • Fecha de publicación del evento
    2014-12-16
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131919
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, tomography, computed, emission - Product Code KPS
  • Causa
    Possibility for system display freeze during ct interventional procedures.
  • Acción
    The firm initiated their recall by certified mail on 11/13/2014.
Retiro De Equipo (Recall) de Device Recall Biograph mCT Flow 404R
  • Tipo de evento
    Recall
  • ID del evento
    69897
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0808-2015
  • Fecha de inicio del evento
    2014-11-13
  • Fecha de publicación del evento
    2014-12-16
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2015-07-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=131920
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, tomography, computed, emission - Product Code KPS
  • Causa
    Possibility for system display freeze during ct interventional procedures.
  • Acción
    The firm initiated their recall by certified mail on 11/13/2014.
  • 1
  • 2
  • 3
  • …
  • Next ›
  • Last »

Acerca de la base de datos

Explore más de 120,000 registros de retiros, alertas y notificaciones de seguridad de dispositivos médicos y sus conexiones con los fabricantes.

  • Preguntas frecuentes
  • Acerca de la base de datos
  • Contáctenos
  • Créditos

Historias en su correo

¿Trabaja en la industria médica? ¿O tiene experiencia con algún dispositivo médico? Nuestra reportería no ha terminado. Queremos oír de usted.

¡Cuéntanos tu historia!

Aviso

Los dispositivos médicos ayudan con el diagnóstico, la prevención y el tratamiento de muchas lesiones y enfermedades. A través de la International Medical Devices Database no estamos sugiriendo que compañías u otras entidades mencionadas en la base de datos hayan sido parte de una conducta ilegal o hayan actuado de manera impropia. Un mismo dispositivo médico puede tener distintos nombres en diferentes países. Esta base de datos no busca proporcionar asesoría médica. Los pacientes deben consultar con sus médicos para determinar si la data contiene información relevante y si la misma tiene implicaciones médicas para ellos.