• Acerca de la base de datos
  • ¿Cómo usar la IMDD?
  • Descargar la data
  • Preguntas frecuentes
  • Créditos
Vista de la lista Vista de las tarjetas
  • Dispositivo 2589
  • Fabricante 31827
  • Evento 2705
  • Implante 0
Retiro De Equipo (Recall) de iSTAT Portable Clinical Analyzer
  • Tipo de evento
    Recall
  • ID del evento
    56832
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0597-2011
  • Fecha de inicio del evento
    2010-09-20
  • Fecha de publicación del evento
    2010-12-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-01-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94604
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Glucose oxidase, glucose - Product Code CGA
  • Causa
    I-stat analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.
  • Acción
    Abbott Point of Care issued Urgent Recall Notice letters dated September 20, 2010 to customers. The letter informs users to of the possibility of the unit becoming hot to touch, to remove the unit if this is the case, not to remove batteries unless instructed to do by APOC and to forward notification of the recall to customers if it has been further distributed. Abbott Point of Care can be contacted at 800 366-8020, option 1, about this action. Recall extended in May 2011 to included distribution dates of 9/20/2010-6/6/2011.
Retiro De Equipo (Recall) de iSTAT Analyzer Military Kit iSTAT 1
  • Tipo de evento
    Recall
  • ID del evento
    56832
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0596-2011
  • Fecha de inicio del evento
    2010-09-20
  • Fecha de publicación del evento
    2010-12-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-01-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94602
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
  • Causa
    I-stat analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.
  • Acción
    Abbott Point of Care issued Urgent Recall Notice letters dated September 20, 2010 to customers. The letter informs users to of the possibility of the unit becoming hot to touch, to remove the unit if this is the case, not to remove batteries unless instructed to do by APOC and to forward notification of the recall to customers if it has been further distributed. Abbott Point of Care can be contacted at 800 366-8020, option 1, about this action. Recall extended in May 2011 to included distribution dates of 9/20/2010-6/6/2011.
Retiro De Equipo (Recall) de iSTAT Abaxis Analyzer
  • Tipo de evento
    Recall
  • ID del evento
    56832
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0595-2011
  • Fecha de inicio del evento
    2010-09-20
  • Fecha de publicación del evento
    2010-12-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-01-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94601
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
  • Causa
    I-stat analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.
  • Acción
    Abbott Point of Care issued Urgent Recall Notice letters dated September 20, 2010 to customers. The letter informs users to of the possibility of the unit becoming hot to touch, to remove the unit if this is the case, not to remove batteries unless instructed to do by APOC and to forward notification of the recall to customers if it has been further distributed. Abbott Point of Care can be contacted at 800 366-8020, option 1, about this action. Recall extended in May 2011 to included distribution dates of 9/20/2010-6/6/2011.
Retiro De Equipo (Recall) de iSTAT Analyzer FUSO Analyzer 300F
  • Tipo de evento
    Recall
  • ID del evento
    56832
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0594-2011
  • Fecha de inicio del evento
    2010-09-20
  • Fecha de publicación del evento
    2010-12-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-01-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94600
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
  • Causa
    I-stat analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.
  • Acción
    Abbott Point of Care issued Urgent Recall Notice letters dated September 20, 2010 to customers. The letter informs users to of the possibility of the unit becoming hot to touch, to remove the unit if this is the case, not to remove batteries unless instructed to do by APOC and to forward notification of the recall to customers if it has been further distributed. Abbott Point of Care can be contacted at 800 366-8020, option 1, about this action. Recall extended in May 2011 to included distribution dates of 9/20/2010-6/6/2011.
Retiro De Equipo (Recall) de iSTAT Analyzer Model 200
  • Tipo de evento
    Recall
  • ID del evento
    56832
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0593-2011
  • Fecha de inicio del evento
    2010-09-20
  • Fecha de publicación del evento
    2010-12-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-01-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94599
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
  • Causa
    I-stat analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.
  • Acción
    Abbott Point of Care issued Urgent Recall Notice letters dated September 20, 2010 to customers. The letter informs users to of the possibility of the unit becoming hot to touch, to remove the unit if this is the case, not to remove batteries unless instructed to do by APOC and to forward notification of the recall to customers if it has been further distributed. Abbott Point of Care can be contacted at 800 366-8020, option 1, about this action. Recall extended in May 2011 to included distribution dates of 9/20/2010-6/6/2011.
Retiro De Equipo (Recall) de ARCHITECT iVancomycin Reagent
  • Tipo de evento
    Recall
  • ID del evento
    56757
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0629-2011
  • Fecha de inicio del evento
    2010-09-13
  • Fecha de publicación del evento
    2010-12-16
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2011-06-22
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94458
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Calibrators, drug specific - Product Code DLJ
  • Causa
    The assay may produce falsely elevated vancomycin results. inaccurate test results may result in inappropriate patient treatment.
  • Acción
    Abbott Laboratories issued a Product Correction letters dated September 13 and 15, 2010 to all customers that received ARCHITECT iVancomycin reagents in the last 12 months. In the September 13, 2010 letter, the accounts were informed of the falsely elevated results and were instructed to continue to follow all sample handling instruction in the package insert and to take the following additional actions to mitigate the risk of obtaining an incorrect result: 1. After following the primary tube manufacturer's recommended centrifugation instructions transfer serum or plasma for all specimens to a microcentrifuge tube for centrifugation at 150,000-300,000 g-minutes, following centrifugation guidelines in the chart in the letter. 2. Prior to testing, transfer the serum or plasma to a sample cup or secondary tube for testing. Care must be taken to transfer only the clarified specimen and not the lipemic material or pellet. The accounts were also requested to complete and fax back to Abbott the enclosed customer reply sheet, acknowledging the receipt and understanding of the recall. On September 15, 2010, Abbott discovered that the calculation for the determination of RCF (xg) given in the September 13, 2010 letter was incorrect. Incorrect Calculation previously provided: Relative Centrifugal Force (RCF) = 0.0000118 x rotational radius in centimeters x RPM2 Correct Calculation within revised letter: Relative Centrifugal Force (RCF) = 0.00001118 x rotational radius in centimeters x RPM2 Abbott stopped distribution of the previous letter and initiated redistribution of the revised letter to all impacted customers, globally, on September 15, 2010. Additionally, all customers which had already received the initial letter were sent a cover letter identifying the error. Customers can contact Abbott at 1-877-422688 about this action.
Retiro De Equipo (Recall) de Architect Tacrolimus Reagent Kit
  • Tipo de evento
    Recall
  • ID del evento
    56719
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0068-2011
  • Fecha de inicio del evento
    2010-09-07
  • Fecha de publicación del evento
    2010-10-15
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2011-05-13
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=94358
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    enzyme immunoassay, tracrolimus - Product Code MLM
  • Causa
    There are reagent barcode read errors when using reagent lot 86599m500 due to the print quality of the barcode.
  • Acción
    Abbott Laboratories sent a "Product Recall Immediate Action Required" letter dated September 7, 2010, to all customers. The letter described product, problem and action to be taken by customers. The customers were instructed to determine if they are currently using and/or have inventory of lot 86599M500; discontinue use of and destroy any remaining inventory; inform and provide copies of letter to other laboratories if appropriate; retain the letter for laboratory records and complete and return the Customer Reply Immediate Action Required form via fax at 1-800-777-0051 or email QAGCO@abbott.com. IMPORTANT NOTE: If you do not currently have replacement material than you may continue to use this reagent lot until your laboratory receives replacement material. Upon receipt of the replacement material, discontinue use of and discard lot 86599M500 and switch to the replacement material. Contact Abbott Customer service for replacement material (1-877-4ABBOTT, Option 1). If you or any of the health care providers you serve have any questions regarding this information, U.S. customers should call Customers Service at 1-877-4ABBOTT. Customers outside of the U.S., please contact your local Customer Service representative.
Retiro De Equipo (Recall) de ARCHITECT i1000SR" Processing Module
  • Tipo de evento
    Recall
  • ID del evento
    55360
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2482-2010
  • Fecha de inicio del evento
    2008-10-21
  • Fecha de publicación del evento
    2010-09-23
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-04-13
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=90666
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Causa
    During manufacturing, two components of the architect i1000sr were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover.
  • Acción
    ABBOTT issued Technical Service Bulletins to Field Service Representatives who then inspected every unit .
Retiro De Equipo (Recall) de Mainline confirms pregnancy test
  • Tipo de evento
    Recall
  • ID del evento
    54131
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1018-2010
  • Fecha de inicio del evento
    2009-10-07
  • Fecha de publicación del evento
    2010-03-08
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2011-02-04
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=87580
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Pregnancy test - Product Code LCX
  • Causa
    Lack of assurance of safety and efficacy: unapproved for marketing in the u.S., non-compliance with cgmp regulations and no stability data to support labeled expirations dates.
  • Acción
    Consignees were notified via recall letters dated October 7, 2009 to cease using and to return the recalled products to Mainline Technology. Direct questions about the recall to Mainline Technology by calling 1-800-541-4467 or 1-734-930-2043.
Retiro De Equipo (Recall) de DIMENSION VISTA CHEMISTRY 3 CALIBRATOR, MODEL KC130
  • Tipo de evento
    Recall
  • ID del evento
    53836
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1535-2010
  • Fecha de inicio del evento
    2009-10-23
  • Fecha de publicación del evento
    2010-05-04
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2010-06-22
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=86626
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Calibrator, multi-analyte mixture - Product Code JIX
  • Causa
    Patient results are depressed.
  • Acción
    An "Urgent Field Safety Notice" dated October 2009 was issued by Siemens to all customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to recalibrate the AMON method (AMON Flex reagent cartridge, K3019) with Vista Chem 3 Claibrator lot 9ED058 using the revised bottle values provided in the notice and to please consult with their laboratory director regarding re-evaluation of previous results. The customers should forward this notification to anyone to whom they may have distributed this product. If you have any technical questions regarding this information, please contact Siemens Technical Solutions Center at 1-800-441-9250.
Retiro De Equipo (Recall) de Whatman Body Fluid Collection Paper 903
  • Tipo de evento
    Recall
  • ID del evento
    53819
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-2384-2010
  • Fecha de inicio del evento
    2009-10-29
  • Fecha de publicación del evento
    2010-09-07
  • Estado del evento
    Open, Classified
  • País del evento
    United States
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=86584
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    tubes, vials, systems, serum separators, blood collection - Product Code JKA
  • Causa
    Neonate newborn screening card demographic and filter portion may be detached.
  • Acción
    The firm, Whatman, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 29, 2009, to the customer. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: inspect the card and ensure the 903 Paper Portion is firmly and completely fixed to the Demographic portio, prior to recording patient demographic details on the card and obtaining a patient sample; discard the card, obtain a new card and inspect as above, if the 903 Paper Portion of the card partially or completely detaches from the Demographic portion of the card; and to please staple the card stub at two points as specified in the illustration, to prevent any opportunity for 903 paper detachment from the demographic portions. If Whatman customers/users have any questions in relation to this customer notification, please contact, either Product Manager at +1 732 885 6570,email: lili.lee@ge.com or Quality Assurance at +1 207 459 7557, email: heidi.kuell@ge.com.
Retiro De Equipo (Recall) de GammaCoat Plasma Renin Activity 125 I RIA Kit
  • Tipo de evento
    Recall
  • ID del evento
    53698
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1914-2010
  • Fecha de inicio del evento
    2009-10-12
  • Fecha de publicación del evento
    2010-06-28
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-04-17
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=86373
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Angiotensin I and Renin Radioimmunoassay - Product Code CIB
  • Causa
    The kit control provided with the gammacoat plasma renin activity kit was labeled with the incorrect range.
  • Acción
    Consignees were Emailed or faxed a "DiaSorin Customer Notification" letter dated October 21, 2009. The letter was addressed to "Dear Valued Customer". The letter described the Product name and the problem and provided the correct range for the Renin Activity Control provided with the kit. The letter also stated that replacement product for any previous failed runs. Customers are to complete the bottom portion of the letter and return it to the firm. Questions or concerns should be directed to DiaSorin Inc., Technical Services at 1-800-328-1482 or 651-439-9710.
Retiro De Equipo (Recall) de Dimension Vista VLYTE Fluids Diluent
  • Tipo de evento
    Recall
  • ID del evento
    53686
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1039-2010
  • Fecha de inicio del evento
    2009-10-22
  • Fecha de publicación del evento
    2010-03-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2010-12-16
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=86323
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Causa
    Results can elevate after sensor replacement. sodium urine results can be falsely elevated 15 mmol/l within 24 hours after installation of a new sensor.
  • Acción
    The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Retiro De Equipo (Recall) de Dimension Vista VLYTE Fluids Standard B/Salt Brige
  • Tipo de evento
    Recall
  • ID del evento
    53686
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1038-2010
  • Fecha de inicio del evento
    2009-10-22
  • Fecha de publicación del evento
    2010-03-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2010-12-16
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=86322
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Causa
    Results can elevate after sensor replacement. sodium urine results can be falsely elevated 15 mmol/l within 24 hours after installation of a new sensor.
  • Acción
    The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Retiro De Equipo (Recall) de Dimension Vista VLYTE Fluids Standard A
  • Tipo de evento
    Recall
  • ID del evento
    53686
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1037-2010
  • Fecha de inicio del evento
    2009-10-22
  • Fecha de publicación del evento
    2010-03-10
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2010-12-16
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=86321
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • Causa
    Results can elevate after sensor replacement. sodium urine results can be falsely elevated 15 mmol/l within 24 hours after installation of a new sensor.
  • Acción
    The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Retiro De Equipo (Recall) de TDx/TDxFLx Gentamicin Wedge Reagent
  • Tipo de evento
    Recall
  • ID del evento
    46146
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0929-2008
  • Fecha de inicio del evento
    2007-10-26
  • Fecha de publicación del evento
    2008-02-26
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-07-30
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=66880
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Fluorescent Immunoassay Gentamicin - Product Code LCQ
  • Causa
    Incorrect results -- through evaluation of stability, abbott have found a decrease in the calibrator a to f span with these reagents lots, which may produce the following results: - printed error code "span less than min span" - controls out of range. when a control is out of range, patient results should not be reported.
  • Acción
    Letters were sent to customers by Federal Express Priority mail on October 26, 2007. Two unique customer letters were sent. One letter was delivered to TDxlTDxFLx customers who received shipment oflots 5ll87Q100 and 51189QI00 and a second letter was delivered to AxSYM customers who received lots 53190QI00 and 53190QI01. Two letters were written because the actions required by the customer differ for the AxSYM and TDxlTDxFLx platforms. TDx/TDxFLx customers " Customer Letter (Product Recall) " Customer Reply Form AxSYM Customers " Customer Letter (Product Correction) " Customer Reply Form The Customers were instructed the following: TDx/TDxFLx Gentamacin Reagents " Discontinue use of and destroy TDx/TDxFLx reagent lots 51187Q 1 00 and 51189QI00. " Document the number of kits discarded for your records. Abbott will provide reimbursement for your destroyed kits. " If customers have forwarded any of the affected lots to another laboratory, to provide the laboratory with a copy of the communication. AxSYM Gentamicin Reagents " Revise the expiration date for lots 53190Ql 00 and 53190QI0l from 4/12/08 to 11/30/2007. " Run each level of control with each AxSYM Gentamicin sample or batch of samples. " If customers have forwarded any of the affected lots to another laboratory, to provide the laboratory with a copy of the communication.
Retiro De Equipo (Recall) de TDx/TDxFLx Gentamicin Reagent
  • Tipo de evento
    Recall
  • ID del evento
    46146
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0928-2008
  • Fecha de inicio del evento
    2007-10-26
  • Fecha de publicación del evento
    2008-02-26
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-07-30
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=66879
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Fluorescent Immunoassay Gentamicin - Product Code LCQ
  • Causa
    Incorrect results -- through evaluation of stability, abbott have found a decrease in the calibrator a to f span with these reagents lots, which may produce the following results: - printed error code "span less than min span" - controls out of range. when a control is out of range, patient results should not be reported.
  • Acción
    Letters were sent to customers by Federal Express Priority mail on October 26, 2007. Two unique customer letters were sent. One letter was delivered to TDxlTDxFLx customers who received shipment oflots 5ll87Q100 and 51189QI00 and a second letter was delivered to AxSYM customers who received lots 53190QI00 and 53190QI01. Two letters were written because the actions required by the customer differ for the AxSYM and TDxlTDxFLx platforms. TDx/TDxFLx customers " Customer Letter (Product Recall) " Customer Reply Form AxSYM Customers " Customer Letter (Product Correction) " Customer Reply Form The Customers were instructed the following: TDx/TDxFLx Gentamacin Reagents " Discontinue use of and destroy TDx/TDxFLx reagent lots 51187Q 1 00 and 51189QI00. " Document the number of kits discarded for your records. Abbott will provide reimbursement for your destroyed kits. " If customers have forwarded any of the affected lots to another laboratory, to provide the laboratory with a copy of the communication. AxSYM Gentamicin Reagents " Revise the expiration date for lots 53190Ql 00 and 53190QI0l from 4/12/08 to 11/30/2007. " Run each level of control with each AxSYM Gentamicin sample or batch of samples. " If customers have forwarded any of the affected lots to another laboratory, to provide the laboratory with a copy of the communication.
Retiro De Equipo (Recall) de Access Ultrasensitive Insulin Assay Kit
  • Tipo de evento
    Recall
  • ID del evento
    45955
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1160-2008
  • Fecha de inicio del evento
    2007-09-19
  • Fecha de publicación del evento
    2008-06-25
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-05-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=66575
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Insulin Assay in-vitro diagnostic kit - Product Code CFP
  • Causa
    False negative results when used to test serum samples (as compared to plasma samples).
  • Acción
    On September 19, 2007, Beckman Coulter mailed to its consignees an Urgent Product Corrective Action Letter informing them that the affected lots of the product were giving false negative results when used to test serum samples (as compared to plasma samples). The letter instructed consignees to discontinue using the affected lots of the product, and to re-test using an EDTA plasma sample if there was any discordance between the results of the serum sample (if it was obtained using an affected lot of the product) as compared with the clinical presentaion of the patient and the results of other diagnostic tests. The letter also instructed consignees to return, within 10 days, the Response Sheet .
Retiro De Equipo (Recall) de Oximetry Ear Clip Sensor
  • Tipo de evento
    Recall
  • ID del evento
    45545
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1092-2008
  • Fecha de inicio del evento
    2007-10-15
  • Fecha de publicación del evento
    2008-04-24
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2009-09-01
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=66529
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    ventilation accessories - Product Code DOA
  • Causa
    Foreign material: glass fragments may present in the plastic bag material used to ship components.
  • Acción
    The recalling firm issued a Recall Notification dated 10/18/07 to its customers. Replacement products were to be sent.
Retiro De Equipo (Recall) de ACCELERATOR Device Manager (ADM)
  • Tipo de evento
    Recall
  • ID del evento
    45873
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0551-2008
  • Fecha de inicio del evento
    2007-10-25
  • Fecha de publicación del evento
    2008-01-25
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2008-04-01
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=66375
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Calculator/Data Processing Module for Clinical Use - Product Code JQP
  • Causa
    Sample/patient mis-identification: two issues identified related to the accelerator decision manager (adm), to include: 1) sample identification (sid) and/or patient identification (pid) numbers that contain more than 12 characters are truncated to the final 12 characters by the adm; 2) for sids using alpha characters that are case sensitive, sids with lower case characters are transmitted by the.
  • Acción
    Firm notified consignees via Product Correction, Immediate Action Required letter(s) on 10/25/07. Consignees were notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Retiro De Equipo (Recall) de Advance Microdraw Blood Glucose Monitoring System
  • Tipo de evento
    Recall
  • ID del evento
    45483
  • Clase de Riesgo del Evento
    Class 3
  • Número del evento
    Z-0458-2008
  • Fecha de inicio del evento
    2007-10-18
  • Fecha de publicación del evento
    2007-12-13
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2008-03-02
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=65563
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Blood Glucose Monitoring System - Product Code CGA
  • Causa
    Mis-labeled/coded strips : meter kits from this lot contain mislabeled 10-count test strip vials. the code number printed on the advance micro-draw 10-count test strip bottle is incorrect. the code number on the vial is 867 when it should be 687. the 687-labeled code chip contains the correct calibration information for the "867" test strips, so the meter will function properly and give accurat.
  • Acción
    Distributors were sent a letter on 10/18/07 asking that product be quarantined. On 10/26/07, an Urgent Voluntary Medical Device Recall letter was sent asking that they return affected product and contact their customers to the retail level.
Retiro De Equipo (Recall) de ACCELERATOR APS System
  • Tipo de evento
    Recall
  • ID del evento
    45429
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0257-2008
  • Fecha de inicio del evento
    2007-10-10
  • Fecha de publicación del evento
    2007-11-24
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2008-03-10
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=65475
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Data processing module for clinical use - Product Code JQP
  • Causa
    Sample problems; five issues identified related to software utilized by the accelerator aps system input/output module, to include: under three specific conditions, the post-aspiration radio-frequency identification (rf id) read of the tube carrier is not performed and the aps work cell may not appropriately generate sample presentation/sample queue errors with aps software version 1.1.1 or earlie.
  • Acción
    Firm notified consignee via letter on 11/10/07. Consignee notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Retiro De Equipo (Recall) de iSTAT G Cartridge
  • Tipo de evento
    Recall
  • ID del evento
    44972
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0375-2008
  • Fecha de inicio del evento
    2007-10-18
  • Fecha de publicación del evento
    2007-12-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2008-07-18
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=64695
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Ion Specific Electrode (Sodium) - Product Code JGS
  • Causa
    Difficult to close -- certain lots of i-stat cartridges are difficult to close, will not close or do not stay closed.
  • Acción
    Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately.
Retiro De Equipo (Recall) de iSTAT EC8 Cartridge
  • Tipo de evento
    Recall
  • ID del evento
    44972
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0374-2008
  • Fecha de inicio del evento
    2007-10-18
  • Fecha de publicación del evento
    2007-12-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2008-07-18
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=64694
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Ion Specific Electrode (Sodium) - Product Code JGS
  • Causa
    Difficult to close -- certain lots of i-stat cartridges are difficult to close, will not close or do not stay closed.
  • Acción
    Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately.
Retiro De Equipo (Recall) de iSTAT CREA Cartridge
  • Tipo de evento
    Recall
  • ID del evento
    44972
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0373-2008
  • Fecha de inicio del evento
    2007-10-18
  • Fecha de publicación del evento
    2007-12-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2008-07-18
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=64693
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Ion Specific Electrode (Sodium) - Product Code JGS
  • Causa
    Difficult to close -- certain lots of i-stat cartridges are difficult to close, will not close or do not stay closed.
  • Acción
    Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately.
  • « First
  • ‹ Prev
  • …
  • 2
  • 3
  • 4
  • 5
  • 6
  • …
  • Next ›
  • Last »

Acerca de la base de datos

Explore más de 120,000 registros de retiros, alertas y notificaciones de seguridad de dispositivos médicos y sus conexiones con los fabricantes.

  • Preguntas frecuentes
  • Acerca de la base de datos
  • Contáctenos
  • Créditos

Historias en su correo

¿Trabaja en la industria médica? ¿O tiene experiencia con algún dispositivo médico? Nuestra reportería no ha terminado. Queremos oír de usted.

¡Cuéntanos tu historia!

Aviso

Los dispositivos médicos ayudan con el diagnóstico, la prevención y el tratamiento de muchas lesiones y enfermedades. A través de la International Medical Devices Database no estamos sugiriendo que compañías u otras entidades mencionadas en la base de datos hayan sido parte de una conducta ilegal o hayan actuado de manera impropia. Un mismo dispositivo médico puede tener distintos nombres en diferentes países. Esta base de datos no busca proporcionar asesoría médica. Los pacientes deben consultar con sus médicos para determinar si la data contiene información relevante y si la misma tiene implicaciones médicas para ellos.

Descargar la data

La International Medical Devices Database está bajo la licencia Open Database License y sus contenidos bajo la licencia Creative Commons Attribution-ShareAlike . Al usar esta data, siempre citar al International Consortium of Investigative Journalists. Puede descargar acá una copia de la base de datos.

Descargar todo (zipped)