Integra Lumbar Drainage Set, REF 910120A, Sterile EO, Rx only, primary packaged in a PVC blister sealed with Tyvek sheet and secondary packaged in to Tyvek pouch. External drainag system is primary packaged in a Tyvek pouch and secondary packaged in to Tyvek pouch. One (1) sealed package of lumbar catheter and External drainage system is finally packaged in a cardboard box. Intended for CSF external drainage
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Algeria India Poland Armenia Indonesia Portugal Australia Iran Austria Iraq Qatar Bangladesh Ireland Republic of Montenegro Barbados Israel Republic of Serbia Belarus Italy Romania Belgium Jamaica Russian Federation Bosnia and Herzegovina Japan Saudi Arabia Botswana Jordan Singapore Bulgaria Kazakhstan Slovakia Cambodia Kenya South Africa Canada Kuwait South Korea Canary Islands Latvia Spain Cayman Islands Lebanon Sri Lanka China Libya Sweden Croatia Lithuania Switzerland Cyprus Macau Syria
Descripción del producto
Medtronic Strata II / Strata NSC Valves (includes Burr Hole and Lumbar Peritoneal. All Catalog Numbers. || Strata-type valves are used in the management of hydrocephalus. The Strata II/Strata NSC valve is a shunt component designed to allow for continuous flow of cerebrospinal fluid (CSF) from the ventricles of the brain to the peritoneal cavity or right atrium of the heart. mechanism. || Strata Valve Magnet Reverse Polarity (Excludes StrataMR). || Catalog No. || 27739 || 27740 || 27786 || 27787 || 27788 || 27789 || 27794 || 27812 || 27814 || 27815 || 27816 || 27817 || 27818 || 27820 || 27821 || 27822 || 27823 || 27824 || 27825 || 27827 || 27828 || 27830 || 27831 || 27832 || 27848 || 27864 || 27867 || 27868 || 27888 || 27903 || 27922 || 27823 || 27924 || 23042 || 27925 || 27926 || 27927 || 27932 || 27933 || 42335 || 42355 || 42365 || 42836 || 42856 || 42866 || 44420 || 44421 || 44430 || 44465 || 46070 || 46075 || 46080 || 46085 || 46635 || 46655 || 46665 || 46836 || 46856 || 46866 || 46871 || 46876 || 46881 || 46886 || 92355 || 92365 || 92856 || 92866 || 96655 || 96665
Product was distributed throughout the US and to Australia, Belgium, Canada, Finland, Germany, Hong Kong, Israel, Italy, Japan, New Zealand, Norway, Poland, Portugal, the Russian Federation, Singapore, Spain, Sweden, Switzerland, Turkey, and the UK.
Descripción del producto
The Medtronic StrataMR adjustable valves and shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to non-invasive adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation. || After surgical implantation, the Medtronic Neurosurgery StrataMR valves provide a non-invasive method to address changing patient needs in the management of hydrocephalus. The valve design enables the physician to adjust the valve pressure/performance level using the Medtronic StrataMR Adjustment Tools. The StrataMR valves are intended to maintain performance level setting under magnetic resonance imaging (MRI) exposure when scanned in accordance with the MRI conditions in the labeling. The StrataMR valves, small and regular, provide the full range of performance levels: 0.5, 1.0, 1.5, 2.0, and 2.5. The currently set performance level is discernible through use of the Indicator Tool or by x-ray (for corresponding settings, refer to figure 6). The dome is designed to allow needle puncture via a 25-gauge or smaller noncoring needle. The StrataMR valve can be flushed in either the distal or proximal direction by percutaneous finger pressure. Occluders are located proximally and distally to the central reservoir on the StrataMR valve regular and only distally on the small model to facilitate selective flushing.
Proclaim 7 Elite, Implantable Pulse Generator, REF 3662, Rx only, STERILE EO || Product Usage: || The Proclaim implantable pulse generator (IPG) is an electronic device designed to be connected to one or more extensions or leads with up to 16 electrodes total. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG can deliver stimulation with a single program or with multiple programs. Each program can provide stimulation to a single anatomical area or to multiple areas. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, || Product Usage: || The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Infinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO || Product Usage: || The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Infinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO || Product Usage: || The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO || Product Usage: || The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Worldwide Distribution - US Nationwide and Puerto Rico and the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Puerto Rico. Distribution internationally to Argentina, Austria, Belgium, Canada, Croatia, Czech Republic, Finland, Germany, Ireland, Israel, Italy, Japan, Lebanon, Netherlands, Norway, Poland, Qatar, Saudi Arabia, United Arab Emirates, and United Kingdom.
Descripción del producto
VNS Therapy¿ AspireHC¿ Generator, VNS Therapy¿ AspireSR¿ Generator. Model No. 105. || Product Usage: || United States Indications for Use: Epilepsy (PMA 970003): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in adults and adolescents over 12 years of age with partial onset seizures that are refractory to antiepileptic medications. Depression (PMA 970003/S050): VNS Therapy is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments. Indications outside the U.S.: Epilepsy (Non-US): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in patients whose epileptic disorder is dominated by partial seizure (with or without secondary generalization) or generalized seizures that are refractory to antiepileptic medications. AspireSR (Seizure Response) features the Automatic Stimulation Mode which is intended for patients who experience seizures that are associated with cardiac rhythm increases known as ictal tachycardia. Depression (Non-US): The VNS Therapy System is indicated for the treatment of chronic or recurrent depression in patients who are experiencing a treatment-resistant or treatment-intolerant depressive episode.
Worldwide Distribution - US Nationwide and Puerto Rico and the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Puerto Rico. Distribution internationally to Argentina, Austria, Belgium, Canada, Croatia, Czech Republic, Finland, Germany, Ireland, Israel, Italy, Japan, Lebanon, Netherlands, Norway, Poland, Qatar, Saudi Arabia, United Arab Emirates, and United Kingdom.
Descripción del producto
VNS Therapy¿ AspireHC¿ Generator, VNS Therapy¿ AspireSR¿ Generator. Model No. 106. || Product Usage: || United States Indications for Use: Epilepsy (PMA 970003): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in adults and adolescents over 12 years of age with partial onset seizures that are refractory to antiepileptic medications. Depression (PMA 970003/S050): VNS Therapy is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments. Indications outside the U.S.: Epilepsy (Non-US): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in patients whose epileptic disorder is dominated by partial seizure (with or without secondary generalization) or generalized seizures that are refractory to antiepileptic medications. AspireSR (Seizure Response) features the Automatic Stimulation Mode which is intended for patients who experience seizures that are associated with cardiac rhythm increases known as ictal tachycardia. Depression (Non-US): The VNS Therapy System is indicated for the treatment of chronic or recurrent depression in patients who are experiencing a treatment-resistant or treatment-intolerant depressive episode
Worldwide Distribution - US including AL, AR, AZ, CA,CO, CT, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN,M MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. Foreign Distribution: India, ANZ, Taiwan, Hong Kong, Canada, China, Japan, Korea, Malaysia, Philippines, Singapore, Thailand, Indonesia, Vietnam, Albania, Algeria, Austria, Belarus, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, India, Iran, Iraq, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, United Kingdom, Argentina, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Mexico, Panama, Peru, Puerto Rico, Uruguay, and Venezuela.
Descripción del producto
DBS Lead Depth Stop contained in Medtronic Deep Brain Stimulation (DBS) lead kits, models 3387/3387S, 3389/3389S, and 3391/3391S, The lead kit contains one DBS lead in addition to multiple accessories used in DBS lead implant procedures. Among these accessories is the DBS lead holder, also referred to as the lead depth stop. || The lead holder affixes to the DBS lead during the lead implant procedure, marking the distance to the lead tip and providing control of DBS lead depth during implant by interfacing with the stereotactic system
Part Number: 10-0011-0500 Serial number/UDI: 8LYB7F2/010542502575039911170920218LYB7F29910-0011-0500 D80C7F2/01054250257503991117092921D80C7F29910-0011-0500
VNS(R) Therapy Programmer, GMDN: 44077, Model 3000 v1.0.2.2 || Intended Use/Indications: United States Indications for Use: Epilepsy : The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in patients 4 years of age and older with partial onset seizures that are refractory to antiepileptic medications. Depression : VNS Therapy is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments.
Lot Number 6210849 Serial Number AVM174.1 Lot Number 6210852 Serial Numbers: AVA560.1, AVA214.1, AVL027.1 Lot Number 6499208 Serial Numbers: AVD426.1, AVA922.1 Lot Number 6263128 Serial Numbers: AVB119.1, AVD400.1, AVA622.1, AVD431.1, AVE633.1, AVK678.1, AVN837.1 Lot Number 6268723 Serial Numbers: AUV770.1, AVW805.1 Lot Number 6278155 Serial Number AVK896.1
Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C. Case Numbers: CSI082114-CT1 CSI081214-TB1 CSI020216-LK1 CSI092016-LD1R CSI100116-BM1A CSI092616-YB1 CSI092616-MM1 CSI112316-YB1 CSI050217-LS1 CSI052217-AF1 CSI062317-GM1 CSI071017-NS1B CSI071417-CM1 CSI102617-DL1 CSI101617-JF1A CSI102617-DL1
Marketing Material: All revisions up to rev. A. IFU: All revisions up to rev. C. Case Numbers: CSI111715-KH1 CSI072814-JB1 CSI082514-MA1 CSI080714-CT1 CSI091914-JW1 CSI101614-TR1 CSI041415-MZ1 CSI100915-DD1 CSI120215-MM1 CSI100915-DD2 CSI011816-TA1 CSI012016-RT1A CSI012716-MP1 CSI021116-CA1 CSI030716-BC1 CSI040716-CK1 CSI060916-DF1 CSI070716-AC1 CSI092116-JA1 CSI110416-DW1 CSI120616-LS1 CSI012417-SD1 CSI062617-HR1
The product is distributed natiowide to 19 different hospitals.
Descripción del producto
EnDura No-React Dural Substitute. Product comes in different sizes (2x10, 4x4, 5x6, 6x10, 10x12, and 12 cm diameters), sterile, stored in 2% benzyl alcohol, and distributed by Integra NeuroSciences, Plainsboro, NJ 08536.
Nationwide and OUS to includeCanada, Europe, Middle East, Africa, Japan, Asia-Pacific, Australia - New Zealand, Latan America
Descripción del producto
Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only); Model 7426 Soletra Neurostimulator; Model 7428 Kinetra Neurostimulator; Models 3387, 3389 DBS Leads; and Models 7482,7495 DBS Extensions. Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. The Medtronic Activa System is an implantable, multiprogrammable system that delivers electrical stimulation to selected areas of the brain. The Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via Extension and Lead. These components comprise the implantable portion of the Activa System.
Nationwide (AZ, CA, FL, ID, IL, IN, KS, KY, MA, MD, MI, MO, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, and WI).
Descripción del producto
Medtronic Model 37711 Restore Neurostimulator. The Restore Neurostimulator is an implanted, rechargeable neurostimulator which provides pain management therapy.
Worldwide, including USA, Puerto Rico, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Israel, Italy, Jordan, Lebanon, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Descripción del producto
Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. The dual program model 7428 Kinetra Neurostimulator generates electrical signals that are transmitted to the brain. These signals are delivered from the neurostimulator to the brain via DBS extension or DBS leads. The neurostimulator consists of electronic circuitry and a battery, which are hermetically sealed in a titanium case. The operation of the neurostimulator is supported by a clinician programmer, a therapy controller, and a control magnet.
Worldwide including USA, Argentina, Australia, Brazil, Czech Republic, Germany, Spain, Finland, France, Great Britain, Greece, Ireland, Netherlands, and New Zealand.
Descripción del producto
Matrix2 Detachable Coils, vascular embolization device. || Catalog Number: 497204SR, UPN: M003497204SR0, || Catalog Number: 494203, UPN: M0034942030, Distributed and Manufactured by Boston Scientific Corporation, Neurovascular Division, Fremont, CA