Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Prosthesis - Product Code KWA
Causa
Instructions for use/surgical technique instructions are inadequate.
Acción
A press release issued on 7/23/08. A letter was sent to implanting surgeons on 7/22/08 advising them of the problem and instructing them to stop implanting these devices until they receive training. Zimmer suspended all marketing and distribution in the U.S. pending completion of user training and revision of the instructions for surgical technique. Zimmer will conduct a removal of all copies of the current surgical technique document and replace it with revised surgical technique document(s). Contact Zimmer, Inc. at 1-866-946-5633 for assistance. Update: A dear surgeon letter was mailed 8/4/08 and provided patient management guidelines. A correction update was sent 8/16/08 with updated surgical technique and instructions for use documents. A dear surgeon letter was mailed 9/25/08 providing additional information regarding the training program.