Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Notas adicionales en la data
Causa
Zimmer biomet is conducting a lot specific medical device recall for various polyethylene implants. the affected products are being removed due to the potential presence of elevated endotoxin levels that exceed the specification limit. the issue was discovered during routine bacterial endotoxin testing (bet). to date, no complaints related to this issue have been received.
Acción
The relevant Zimmer Biomet sales representative will remove the affected product from each affected facility. There are no specific patient monitoring instructions that are recommended beyond the existing surgical follow-up protocol.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Notas adicionales en la data
Causa
Baxter healthcare is issuing a recall of actifuse abx and actifuse mis system products with expiry before 29 july 2017 due to the possibility that the products may have endotoxin levels above specification criteria. in surgical procedures where there is device contact with the cerebrospinal fluid through a dural opening (iatrogenic injury), the use of a medical device with increased endotoxin levels may augment the typical inflammatory reaction to surgery and contribute to adverse health consequences. baxter has not received product-related adverse event reports that can be linked to cerebrospinal fluid exposure to increased levels of endotoxins.
Acción
Users are asked to inspect their stock and to remove affected product from their facility and to contact Baxter Customer service to arrange replacements. This action has been closed-out on 05/09/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Notas adicionales en la data
Causa
Medtronic has become aware of instances where spine clamps have been damaged when forced open beyond their intended limits during use and will subsequently not open once attached to a patient’s spinous process. when the spine clamp is forced open beyond its intended limits, a component (captive washer) may break off. the washer could then be inadvertently left behind in a patient’s body if the breakage occurs during the procedure. if the washer is missing from the device, the spine clamp cannot be re-opened after placement on the spinous process.Medtronic has received six reports where unintended removal of spinous process occurred when attempting to detach the spine clamp. the unintended removal of spinous process can lead to damage of adjacent vertebra and cause premature degradation.
Acción
Medtronic is providing an interim workaround for users via the Customer Letter. A new version of the clamps with a design mitigation will be distributed as a permanent correction once available.
Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
spinal bone tamp - Product Code HRX
Causa
Product not sterile: 10 units were released for distribution without undergoing sterilization process.
Acción
Sales representatives were contacted via e-mail on February 4, 2009, and instructed to contact hospitals and inform them of the issue and sequester affected product. They were further instructed to immediately travel to the hospitals and obtain affected product for return to Medtronic LLC. A Customer Communication Notice URGENT NOTICE: DEVICE REMOVAL dated February 9, 2009, was sent to the two consignees.