Lebanese data is current through July 2018. All of the data comes from the Ministry of Public Health (Republic of Lebanon), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Lebanon.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Notas adicionales en la data
Causa
Invacare has recently identified a potential pinch point affecting the wiring of the listed power wheel chairs. there is a possibility that if the slack in the wires is not routed and secured correctly, and with flexing of the back pan under the user’s weight, the headrest knob or clamp bracket may pinch, damage or cut the wire creating a hazard. smoking, sparking, burning or fire may occur if any wire in this area is repetitively pinched, at the same location, over a period of time. not all impacted chairs are expected to have slack in the wiring that is not routed and secured correctly. invacare has received one complaint and there have not been any reported events involving injury associated with this issue.
Acción
Dealers are advised to identify wheelchair owners and notify them of this issue and proposed actions using the consumer letters provided by Invacare. Dealers are also advised to contact their respective customers to schedule an appointment to inspect the potentially impacted power wheelchair as soon as possible. A visual inspection of the wiring shall be conducted for each impacted chair and where required the wiring will be (corrected) routed and secured in accordance with the Power Wheelchair Wiring Guide provided to dealers. For more details, please see http://www.tga.gov.au/alert/invacare-tdx-and-storm-series-power-wheelchairs . This action has been closed-out on 18/08/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Notas adicionales en la data
Causa
A ge healthcare internal quality inspection has found that the power cord connecting directly to the t2100 and t2000 treadmills may not have been assembled according to specifications. if the power cord was improperly assembled and a separate secondary electrical fault condition exists (e.G., a frayed extension or power cord touching the treadmill chassis), this could possibly result in an electrical shock to the patient or operator. there have been no reported incidences of the treadmill power cord leading to an electrical shock of a patient or operator at this time.
Acción
GE Healthcare is instructing customers to continue to perform the routine maintenance specified in the T2100 T2000 Service Manuals. This includes the leakage tests performed after each monthly internal cleaning, as specified in the T2100 and T2000 Service Manuals. A GE Healthcare Service Representative will contact customers to arrange a correction. This action has been closed-out on 15/08/2016.
Data from Andorra is current through November 2018. All of the data comes from Health Andorra, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Andorra.
Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland.
Notas adicionales en la data
Advice regarding a device removal. 3rd Party Publications
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Bed, patient rotation, powered - Product Code IKZ
Causa
Product complaints indicating that while attempting to resolve an alarm or error condition, operator was unable to return product to the prone position for patient.
Acción
KCI USA, Inc. sent an URGENT - Medical Device Safety Alert letter, dated September 8, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the cusotmer. Customers were instructed to have a practiced back-up plan, such as a plan for the manual proning of patients in the event that clinical circumstances dictate the need for more immediate prone therapy.
For questions, customers in the US should call KCI's Technical Support Center at 1-800-275-4524, option 5. For customers outside the US they should contact their local sales representative.
Customers should contact their accounts with affected product and advise them of the safety notice.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chair, with casters - Product Code INM
Causa
Winco mfg., llc, ocala, florida is recalling their care cliner. due to malfunctioning axles which become lose from the caster assembly. this chair is a multi position recliner with a steel frame used primarily in clinics, hospitals and treatment centers for recovery, dialysis, oncology, examination, infusion. recalled chairs include the following: care cliner, xl care cliner, drop arm care cliner.
Acción
Winco sent an Urgent Device Recall letter dated October 1, 2010, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. Replacement parts were sent out to the consignees and customers were asked to dispose of the defective casters. Any consignee not reponding by November 15, 2010, would receive an additional letter.
The firm placed a follow up phone call to the non responding consignees after two weeks.
Customers were advised to call 800-237-3377 or email customer service at customerservice@wincomfg.com with questions or concerns.
For questions or concerns regarding this recall call 352-854-2929 x 110.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Powered wheelchair - Product Code ITI
Causa
For some invacare chairs, the stability lock feature may not be engaging properly or consistently. if the stability lock feature does not engage properly, the wheelchair may do one or more of the following: (1) veer to one side; (2) rock forward onto its front riggings; (3) drive in an unintended circular motion; and/or (4) fall forward or to one side. in these cases there is a risk to injury for.
Acción
On 9/15/2009 the firm sent recall notification letter to their customers instructing them to immediately contact the dealer to have the stability lock inspected and/or adjusted. If the dealer is unavailable or the repair will take longer than 60 days, they are to contact the firm and arrange for an inspection and/or adjustment. Invacare requested contact names to whom it can send additional information. Invacare plans to mail, via FEDEX, information regarding the recall to each consignee. Initial consignees received a response card and instructions to forward cards to any customer to whom they have distributed the product. The products are to be inspected at the cylinder label and if the label does not have a long alpha-numeric string demonstrated in the label, the cylinders on the chair have to be replaced. Questions are directed to Recalls@Invacare.com or to 800-333-6900.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
System, Communication, Powered - Product Code ILQ
Causa
Pendant personal help button neck cord may not break away and if not will present a potential choking risk.
Acción
Philips Lifeline initiated Safety Notification letters on 9/21/09 to users and advised of the potential choking risk of the Pendant Personal Help Button The firm recommends that users consult their health care providers to determine which style of emergency button, including those that are worn on the wrist, is most beneficial for them. A postage-paid reply envelope had been provided. Questions should be addressed to Philips at 1-877-221-8756 (Monday Friday, 8 a.m. 8 p.m. EST).
Firm issued Press on 9/11/09. FDA Press issued 9/22/09
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Device, prosthesis alignment - Product Code IQO
Causa
The bubble in a vial of the pelvic alignment level (pal) was wider than the level marking. upon opening the pal and removing the vial, it was discovered that the vial was leaking.
Acción
Stryker Orthopaedics issued an "Important Product Removal" letter dated September 11, 2007 via Federal Express informing consignees of the affected devices. The firm requested that users identify and return any affected product and account for product by returning a Product Accountability Form via fax.
For further information, contact Stryker Orthopaedics at 1-201-831-5825.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Otto bock health care lp has discovered that in some cases, the securing screws of the joint axes may not have been properly mounted. this can cause the axes to gradually move out of the joint and result in a risk of falling for the user.
Acción
Consignees were sent on 10/28/08 a Otto Bock "Urgent: Correction Notice" letter dated October 27, 2008. The letter described the product and problem. It provided instructions if the product was in stock and instructed the consignee to contact the patients (a patient letter was provided) and to complete and return the reply card. Consignees were to contact Otto Bock at their Customer Service to request a return authorization.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Transfer Aid - Product Code IKX
Causa
Misassembled gait (transfer aid) belt.
Acción
Alimed notified accounts via U.S. Postal Service (Return Receipt Requested) with a Recall Notice dated 10/30/2008 and instructs users to review their records and notify all users to whom the product may have been issued. If any of the product has been issued outside of their facility, customers are to forward the correspondence to them in it's entirety. Customers are to complete the enclosed form for return of their product (new or used) and the issuance of replacement product or credit.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Transfer Aid - Product Code IKX
Causa
Misassembled gait (transfer aid) belt.
Acción
Alimed notified accounts via U.S. Postal Service (Return Receipt Requested) with a Recall Notice dated 10/30/2008 and instructs users to review their records and notify all users to whom the product may have been issued. If any of the product has been issued outside of their facility, customers are to forward the correspondence to them in it's entirety. Customers are to complete the enclosed form for return of their product (new or used) and the issuance of replacement product or credit.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Toddler Chair - Product Code INN
Causa
Some hip straps were assembled incorrectly by threading the buckle on the strap upside down. if a child were left unattended in spite of the warnings on the product and in the manual, a defective hip strap could loosen, allowing the child to slip down in the chair and this could create a choking hazard.
Acción
On 9/12/08, Rifton Equipment began notifying the customers by phone using a phone script. This phone call notified the customers of the recall and also served as an effectiveness check that they were notified. The customers were instructed to remove the product from use until the replacement kit is received and the old strap is returned. The replacement kit contains instructions with ID codes and a place for the customer to verify that they have made the replacement as instructed. It also contains the required number of new hip straps. A verification check was sent with the replacement straps, which the customer will sign and return to the firm verifying that they have replaced the straps for the affected serial numbers. Rifton Equipment is providing the customer with return shipping labels which were sent by UPS 2nd day air. The replacement straps use black thread instead of tan thread for the sewing on the ends of the straps so that the firm can be sure that the returned straps are not the new replacement straps.
For questions or additional information, contact Rifton Equipment at 1-800-571-8198.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Some units of fluro-ethyl containing the defective valve which could malfunction and spray refrigerant out from the side of the valve in addition to or instead of spraying in the normal inverted position from the product's actuator. the product can also leak coolant from the side of the valve onto the fingers of the user.
Acción
A recall notification letter (Recall Notice) (dated 9/3/2008) was issued on 9/16/2008 to customers. The notice instructs customers to check their inventory for any remaining recall product and if found, return the product to the Gebauer Company. The letter also advises distributors to notify their customers and follow instructions as noted above, i.e. customers inventory their stock for recalled product and if found, return to the Gauber Company. For questions or additional information, contact Gebauer Company at 216-581-3030 (ext. 120) or 800-321-9348.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Computed Radiography System - Product Code MBQ
Causa
Users may experience one or more of four (4) problems; 1. image quality problem with full leg full spine exams. 2. unable to print 2 or more images on 1 film. 3. possible image loss when printing. 4. study date and time not showing up on imexius web.
Acción
A Service Bulletin to address the issues was created on 09/01/2006. Customers were notified of the plans to upgrade via telephone and email on 10/13/2006.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Tube Clamp adapters - Product Code ISH
Causa
Device fracturing: recent design and process changes leave the device susceptible to fracturing, which could cause the prosthesis to fail.
Acción
On October 6, 2006, SPS faxed a recall notification to the 187 customers who had purchased one or more BL120M Tube Clamp Adapters from SPS between August 29 and September 20, 2006. On October 9, SPS contacted those customers that had purchased five or more Adapters via telephone to confirm they had received the faxed recall notification. On October 10, SPS mailed another copy of the recall notification, via certified mail, return receipt requested to all of its customers that had purchased one or more BL120M Tube Clamp Adapters. The customers were told to send the BL120M Adapter back to SPS following replacement.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Component, External, Limb, Ankle/Foot - Product Code ISH
Causa
After one year devices assembled with a new type of o-rings could exhibit enough wear to allow a gradual oil leak which would increase play in the ankle sub-assembly of the foot.
Acción
Telephone communication followed by notification letter mailed to end users (patients) and prosthetics workshops. A pre-paid package to return faulty components to Ossur included with mailed letters. New replacement feet will be shipped to customers.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wheelchair, Powered - Product Code ITI
Causa
When the powered seating system is tilted completely in the rearward position, a potential pinch point is created between the seat and the seat frame.
Acción
The recalling firm sent letters to all consignees.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Bed, Patient Rotation, Powered - Product Code IKZ
Causa
Design control/validation deficiencies.
Acción
Letter dated 10/25/2004 to Stryker Medical, own-label distributor, with instructions to remove patients from the system and return units to Gaymar.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wheelchair, Mechanical - Product Code IOR
Causa
Under certain conditions the tilt mechanism of the strollers may not function properly allowing the child to fall out if the seat belt is not fastened.
Acción
Recall letters dated October 23, 2003 were sent to the consignees on 10/29/03 and 10/30/03. The letters requested that the consignees contact the firm''s Customer Service Department to report the affected units they have. The letter states that the Customer Service Department will send the consignees the necessary field modification kits and instructions to correct the problem in the units they have and have located at their customers.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Otto bock healthcare products gmbh has identified a design issue through extended durability
testing where the pylon clamping mechanism has the potential to cause damage to the
pylon. when used longer than 1.45 million cycles (which represents approximately 2.3
years of use of the knee joint under heavy usage conditions), and the pylon clamp is overtightened
according to the labeled torque specification, and the patient weighs at most
125kg it may be remote that the pylon breaks, and causing the patient to fall.
there have been no failures or complaints reported directly attributed to the failure mode
collapsed pylon.
in the us market and in canada, labeling requires a condition based service at three years.
in the rest of the world, a mandatory two year service interval is specified. only the devices
in distribution in the united states and canada are affected by this action.
Acción
On September 14, 2017 a letter was sent to all their consignees via 2nd day signature receipt required to customers instructing them to contact their patients and send in their devices for service within the first 2.3 years of use. The pylon clamp will be replaced at no charge. The pylon will also be inspected and replaced if necessary. The customer will be requested to complete a reply form indicating they have received the notice and intend to contact their patients to arrange for the requested service.
Customer Service: (800) 328-3458 Option 1
Loaner Service: (800) 328-3458 Option 3, then Option 1
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Otto bock healthcare products gmbh has identified a design issue through extended durability
testing where the pylon clamping mechanism has the potential to cause damage to the
pylon. when used longer than 1.45 million cycles (which represents approximately 2.3
years of use of the knee joint under heavy usage conditions), and the pylon clamp is overtightened
according to the labeled torque specification, and the patient weighs at most
125kg it may be remote that the pylon breaks, and causing the patient to fall.
there have been no failures or complaints reported directly attributed to the failure mode
collapsed pylon.
in the us market and in canada, labeling requires a condition based service at three years.
in the rest of the world, a mandatory two year service interval is specified. only the devices
in distribution in the united states and canada are affected by this action.
Acción
On September 14, 2017 a letter was sent to all their consignees via 2nd day signature receipt required to customers instructing them to contact their patients and send in their devices for service within the first 2.3 years of use. The pylon clamp will be replaced at no charge. The pylon will also be inspected and replaced if necessary. The customer will be requested to complete a reply form indicating they have received the notice and intend to contact their patients to arrange for the requested service.
Customer Service: (800) 328-3458 Option 1
Loaner Service: (800) 328-3458 Option 3, then Option 1
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wheelchair, powered - Product Code ITI
Causa
There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
Acción
The firm initiated the recall by letter on 09/01/2017. The distributor/dealers were instructed as follows:
LOG INTO CORRECTION WEB PORTAL
Immediately log into www.permobilfsa.com. Log in with the following ID and Password.
CEASE DISTRIBUTION OF EXISTING INVENTORY
Please examine your inventory for the serial numbers assigned to you, cease distribution, and quarantine those chairs.
IMMEDIATELY PERFORM INSPECTION
If you have distributed these serial numbers to end users, you must immediately contact them to perform a visual inspection of the top plate assembly looking for obvious signs of cracking.
If the top plate shows no signs of cracking the wheelchair may continue to be used in accordance with the owners manual until the correction is performed.
If the top plate does show signs of cracking please inform the user that the chair cannot be used until the correction is performed. Next, log into the web portal to obtain further instructions and assistance. In these cases which are expected to be rare, we will assist with a temporary solution to keep the client mobile during the time between inspection and correction.
PERFORMING THE CORRECTION
You will be notified when repair kits are available and the process to follow to obtain the proper kit for the serial number you are correcting.
The web portal will include a template letter to send to end users explaining the hazard and the need to cooperate with this correction effort. Correction instructions and related supporting materials will also be available. We are also providing the end user with a certificate for a free light kit towards the purchase of a new Permobil power chair.
Please contact your Permobil sales representative if you have any questions regarding this correction.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wheelchair, powered - Product Code ITI
Causa
There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
Acción
The firm initiated the recall by letter on 09/01/2017. The distributor/dealers were instructed as follows:
LOG INTO CORRECTION WEB PORTAL
Immediately log into www.permobilfsa.com. Log in with the following ID and Password.
CEASE DISTRIBUTION OF EXISTING INVENTORY
Please examine your inventory for the serial numbers assigned to you, cease distribution, and quarantine those chairs.
IMMEDIATELY PERFORM INSPECTION
If you have distributed these serial numbers to end users, you must immediately contact them to perform a visual inspection of the top plate assembly looking for obvious signs of cracking.
If the top plate shows no signs of cracking the wheelchair may continue to be used in accordance with the owners manual until the correction is performed.
If the top plate does show signs of cracking please inform the user that the chair cannot be used until the correction is performed. Next, log into the web portal to obtain further instructions and assistance. In these cases which are expected to be rare, we will assist with a temporary solution to keep the client mobile during the time between inspection and correction.
PERFORMING THE CORRECTION
You will be notified when repair kits are available and the process to follow to obtain the proper kit for the serial number you are correcting.
The web portal will include a template letter to send to end users explaining the hazard and the need to cooperate with this correction effort. Correction instructions and related supporting materials will also be available. We are also providing the end user with a certificate for a free light kit towards the purchase of a new Permobil power chair.
Please contact your Permobil sales representative if you have any questions regarding this correction.