Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Catheter shaft breaks. including cases requiring the use of a snare or other intervention to retrieve the broken shaft from the patient. no serious injuries have been reported.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Incorrect configuration setting cause incompatibility between the onetouch ping insulin pump and ezmanager max diabetes management software affecting the ability to download view and print pump information.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Device was shipped with the incorrect "instructions for use" (ifu) in the package. this error only affectes the "indications for use" section.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Lot 74264un11 is giving reduced rlu values compared to other lots so there is the potential for incorrect patient results for troponin which can lead to incorrect management of patients with suspected acute coronary syndrome.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Some issues have been reported when firmware version 3.13 was installed on systems equipped with the old transient recorder (tr) board (ref. 6200350). extenisve testing has proven that firmware 3.13 may potentially cause an instrument to freeze or hang during acquisition when installed with the old tr board with a risk repeated results for different spots.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential that the 11.4 cm introducer needle within certain lots of model 8731sc intrathecal catheter kits may exceed usp requirements for bacterial endotoxin.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
Notas adicionales en la data
Causa
Continuous recycling and/or rebooting of plum a+ devices may occur when the "backlight intensity" and/or "display contrast" settings for the lcd display have been adjusted from the original default setting.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential of imprecision of the online dat cannabinoids ii (thc ii) that may increase the possibility to obtain results under the cut-off for patient samples whose concentration is very close to the cut-off.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Elimination of software bug which allows for improvement in software stability and for new functions.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Biomerieux has observed an increase in the failure rate of the vitek ms plus-chca matrix when used as the negative control (matrix run alone on the spot).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Diagnostica stago was aware that sta system control n+p factor viii results were not within the assigned acceptance range.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
La stabilite de la bilirubine totale des lots de liquichek pediatric control differe des valeurs publiees dans la notive explicative.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Beckman coulter has received reports regarding ampicillin and penicillin false resistance with clinical isolates and out-of-range (high) quality control (qc) results on the microstrep plus 1 (mstrp+1) and microstrep plus 2 (mstrp+2) panel lots included in this recall. beckman coulter investigation has confirmed a manufacturing error leading to antimicrobic degradation and the potential for elevated or false resistant results for some antimicrobics (e.G. ampicillin and penicillin) in a portion of the panels within these lots.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Fracture of the nitinol frame within the aneurx stent graft following implantation (time of occurrence unknown).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
While performing a service call it was identified that a somatom definition as was missing the laser windows on the gantry front cover. it was identified that due to the gantry cover opening there is a risk for severe injury of the patient if he/she is to use the opening for stabilization or for support. further investigation revealed that the somatom definition as and the somatom definition flash were affected.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Potential compromise to sterility in breastpump package in rigid trays with tyvek lids.It is due to pinholes in the rigid tray most likely created during the tray forming process.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Increased frequency of control values recovering out high of the defined range when run on the liaison xl analyzer.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The use of allura xper systems regarding live images and still images can lead to still images being interpreted as live images.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The specification for "high dose hook effect" for the cysc2 on all roche systems has been revised and will be reduced from 20 mg/l to 12 mg/l.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Suite a une etude interne il a ete determine que le produit une fois ouvert restera stable pendant 5 jours et non pas 16 jours.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The package insert provided with the listed kit lots of the monolisa anti-hbs eia contains a typographical error. following this erronous instruction would result in a 1:6 dilution and invalidate the assay. notice: recall initiated and closed in 2008.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
When all plans are loaded on the device if the beam plan is edited accessory information in the lantis can be potentially overwritten or removed could result in mistreatment on another linear accelerator.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Maneuvers could reactivate functionalities of the device (saf-3000/saf-4000) that are not permitted by canadian regulations.