The cartridge leak indicator of the STERRAD® 100NX® system cartridges, Ref .: 10144, manufactured by Advanced Sterilization Products (ASP), USA, is only effective for 12 of the 15 months of the cartridge's validity period.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Cessation of the sterilization of certain models of Karl Storz endoscopes, in Sterrad® systems, manufactured by Advanced Sterilization Products (ASP), because they incorporate a material in the light that is not suitable for treatment in these systems.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that the vacuum pump oil of certain STERRAD® 100S Sterilization Systems, part number 10101, manufactured by Advanced Sterilization Products (ASP), USA, does not meet the manufacturing specifications.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Introduction to the sterilization process of a checklist to be used with the "STERRAD® enhancers and adapters", manufactured by Advanced Sterilization Products (ASP), USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possible malfunction of the leakage indicator of the cartridge packaging system for STERRAD® Sterilization Systems, manufactured by Advanced Sterilization Products (ASP), USA, when exposed to hydrogen peroxide.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Precautionary cessation of the use of lot 12C1203 of the product "AMO Endosol®", Ref .: 15020, manufactured by AMO Groningen BV, The Netherlands, as five cases of Previous Segment Toxic Syndrome were reported.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal of the precautionary cessation of the use of lot 12C1203 of the product "Irrigation solution AMO Endosol®", Ref. 15020, manufactured by AMO Groningen BV, Holland, since the suspicion of contamination of said batch was ruled out.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
The hydrogen peroxide plasma gas technology of the STERRAD® 100S System should not be used for the sterilization of Anspach products, manufactured by Anspach Effort Inc, USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Safety warning related to the navigation systems CARTO® 3 EP, manufactured by Biosense Webster, Israel, when the intracardiac ultrasound images disappear if the power supply to the system is interrupted.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that with the navigation systems CARTO® 3 EP, manufactured by Biosense Webster, Israel, leakage of current occurs that could induce a ventricular arrhythmia to the patient.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal from the Thermocool® SmartTouch ™ catheter market, manufactured by Biosense Webster Inc., USA, due to the possibility of interruption of the flow of the irrigation liquid.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Identification of an increase in the frequency of adverse events associated with cardiac perforation / tamponade and atrioesophageal fistulas with the ThermoCool® SF NAV catheter family, manufactured by Biosense Webster Inc., USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility of a curvature or rupture in different body locations of the THERMOCOOL SMARTTOUCH catheters, manufactured by Biosense Webster Inc., USA, when a manual pre-bending of the catheter is performed or 8Fr sheaths are used
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Withdrawal from the market of certain batches of the Irrigated Circular Catheter nMARQ, with catalog number D132214, manufactured by Biosense Webster Inc., USA, due to a malfunction of the thermocouples
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Withdrawal from the market of certain references and batches of the PENTARAY® NAV catheter, manufactured by Biosense Webster, Inc., Germany
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Extension of the recall to all batches of the Pentaray® NAV and Pentaray NAV ECO catheter, manufactured by Biosense Webster, Inc., Germany
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Risk of entrapment of PENTARAY® NAV and PENTARAY® NAV eco catheter arms, manufactured by Biosense Webster, Inc., Germany, when used in patients with prosthetic valves.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility of ablation radio frequency interference through certain interconnection cable configurations in certain CARTO® 3 Systems, manufactured by Biosense Webster, Ltd, Israel.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that when combining FJORD cemented femoral stems, manufactured by DePuy France, with metal-metal joints, excessive corrosion of the polished surfaces that are inside the cement mantle will occur.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Identification of different problems related to certain references of the ATTUNE® Knee System Instrument, manufactured by Depuy International Ltd, United Kingdom.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Extension of the market recall to certain codes and lots of the LUPINE® suture anchor system, manufactured by DePuy Mitek Inc., USA, due to the possibility that they may contain a suture not properly assembled.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Removal from the market of all codes and batches of the suture anchors of the SPIRALOK product, and discontinuation of this product line, manufactured by Depuy Mitek, USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possible breakage of the tabs of the connecting end of the HP MBT keel knee instrument, used for the preparation of the tibial canal during total knee arthroplasty, manufactured by Depuy Orthopedic Drive, USA.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility of holes being formed in the containers of the rotating hinge S-ROM Noiles de DePuy (femur with pin), manufactured by DePuy Orthopedic Inc, United States.
The data from Spain is current through June 2018. All of the data comes from Agencia Española de Medicamentos y Productos Sanitarios, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Spain.
Notas adicionales en la data
Acción
Possibility that the "GLOBAL APG + 2.5 mm dividable guide", product code 223000019, manufactured by DePuy Orthopedic Inc., USA, may break and remain in the patient.