Retiro De Equipo (Recall) de Siemens Artis Zeego X-ray System with Software Artis
  • Tipo de evento
    Recall
  • ID del evento
    14315
  • Fecha de inicio del evento
    2013-03-08
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • Causa
    The manufacturer advises that under certain preconditions, executing roadmap ogp (organ program) in the dsa overlay mode may lead to imprecise registration of the dsa mask image with the roadmap image displayed on the live monitor.
  • Acción
    Software to be upgraded
Retiro De Equipo (Recall) de Stryker Switch Point Infinity 3 (SPI3)
  • Tipo de evento
    Recall
  • ID del evento
    14314
  • Fecha de inicio del evento
    2013-03-07
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Stryker New Zealand Ltd, 515 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
  • Causa
    Stryker has become aware that for sources utilizing s-video and composite inputs, pal analogue video signals appear with jagged edges on all diagonal lines resulting in poor video quality.
  • Acción
    Software to be upgraded
Retiro De Equipo (Recall) de Kerr Extrude Wash dental impression material
  • Tipo de evento
    Recall
  • ID del evento
    13227
  • Fecha de inicio del evento
    2012-10-02
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Henry Schein New Zealand, 23 William Pickering Drive, Albany, North Shore City, Auckland
  • Causa
    Instructions for use contain the incorrect timing parameters for this product.
  • Acción
    Instructions for use to be updated
Retiro De Equipo (Recall) de Invacare Power Wheelchairs
  • Tipo de evento
    Recall
  • ID del evento
    13210
  • Fecha de inicio del evento
    2012-07-13
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Invacare NZ Ltd, 4 Westfield Place, Mt Wellington, AUCKLAND
  • Causa
    Post fuse assemblies on certain modela of invacare power wheelchairs, which could potentially be defective in a manner that would result in the fuse not operating as intended (blowing) in the event of future damage or faults within the main wiring harness., while each battery is equipped with a fuse that alone would provide the intended safety function , there is a potential risk of fire and serious injury should both fuses be defective.
  • Acción
    Product to be modified
Retiro De Equipo (Recall) de Abbott i-STAT cTnl Cartridge
  • Tipo de evento
    Recall
  • ID del evento
    13199
  • Fecha de inicio del evento
    2012-07-25
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Ebos Group Ltd, 14-18 Lovell Court, Rosedale, Auckland 0632
  • Causa
    Abbott point of care has removed the recommendation to use blood collection devices that are at least half (1/2) full from i-stat system user documentation. apoc recommends that samples are obtained from blood collection devices filled to their stated volume regardless of cartridge or sample type.
  • Acción
    Instructions for use to be updated
Retiro De Equipo (Recall) de Linvatec PressFT Polyethylene HI-FI Sutures
  • Tipo de evento
    Recall
  • ID del evento
    13195
  • Fecha de inicio del evento
    2012-08-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: REM Systems Limited, 69 Elizabeth Knox Place, Glen Innes, AUCKLAND
  • Causa
    Potential for the distal tip of the driver to become separated from the driver shaft after successful insertion of the implant.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Ormco AEZ Mini Distal Cutter w/ Hold Long
  • Tipo de evento
    Recall
  • ID del evento
    13194
  • Fecha de inicio del evento
    2012-07-30
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Law and Accounting Professional Trustees Limited, 336, Jackson Street, Petone
  • Causa
    Aez mini distal custter labeled as long-handle cutters actually contain short-handle cutters.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Abbott RealTime High Risk HPV Amplification Reagent Kit
  • Tipo de evento
    Recall
  • ID del evento
    13193
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Abbott Diagnostics Division, Ground Floor, Building D, 4 Pacific Rise, Mt Wellington, AUCKLAND 1640
  • Causa
    The manufacturer has received reports of barcodes on reagent packs being unable to be scanned due to the barcodes being diffuse and/or having white spots within the barcode which prevents them from being read consistently by the m2000sp instrument.
  • Acción
    Manufacturer to issue advice regarding use
Retiro De Equipo (Recall) de Philips Brilliance series 6, 16, 40, 64 CT system
  • Tipo de evento
    Recall
  • ID del evento
    13192
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    In the ct systems above, sixteen bolts secure j-brackets which support the carbon table top to the couch subframe. philips healthcare received two complaints stating that the bolt heads of several of these sixteen bolts had broken off., a risk of serious injury (bone fracture, head injury, concussion) was identified in the event that all sixteen (16) bolts are broken. if this occurs, and the carbon table top is extended, the table top may tilt or tip over, or it may result in unintended pivotal motion of the table. if the patient is not secured to the table, there is a possibility the patient may slide off, which may cause serious injury. if less than sixteen (16) of the j-bracket bolts fail, the carbon table top will remain attached to the couch subframe.
  • Acción
    Product to be modified
Retiro De Equipo (Recall) de Philips CT Big Bore CT system
  • Tipo de evento
    Recall
  • ID del evento
    13192
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    In the ct systems above, sixteen bolts secure j-brackets which support the carbon table top to the couch subframe. philips healthcare received two complaints stating that the bolt heads of several of these sixteen bolts had broken off., a risk of serious injury (bone fracture, head injury, concussion) was identified in the event that all sixteen (16) bolts are broken. if this occurs, and the carbon table top is extended, the table top may tilt or tip over, or it may result in unintended pivotal motion of the table. if the patient is not secured to the table, there is a possibility the patient may slide off, which may cause serious injury. if less than sixteen (16) of the j-bracket bolts fail, the carbon table top will remain attached to the couch subframe.
  • Acción
    Product to be modified
Retiro De Equipo (Recall) de Philips iCT and iCT SP CT systems
  • Tipo de evento
    Recall
  • ID del evento
    13192
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    In the ct systems above, sixteen bolts secure j-brackets which support the carbon table top to the couch subframe. philips healthcare received two complaints stating that the bolt heads of several of these sixteen bolts had broken off., a risk of serious injury (bone fracture, head injury, concussion) was identified in the event that all sixteen (16) bolts are broken. if this occurs, and the carbon table top is extended, the table top may tilt or tip over, or it may result in unintended pivotal motion of the table. if the patient is not secured to the table, there is a possibility the patient may slide off, which may cause serious injury. if less than sixteen (16) of the j-bracket bolts fail, the carbon table top will remain attached to the couch subframe.
  • Acción
    Product to be modified
Retiro De Equipo (Recall) de Covidien Shiley Tracheostomy Tube, Cuffed with Inner Cannula
  • Tipo de evento
    Recall
  • ID del evento
    13189
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Covidien NZ Ltd, Level 3, Building 5, Central Park Corporate Centre, 666 Great South Road, Penrose, AUCKLAND 1051
  • Causa
    The manufacturer has received reports of certain size 8 shiley™ reusable cannula, cuffed tracheostomy tubes that have had volume leakage and/or disconnection between the inner and outer cannulae. these events were typically observed during mechanical ventilation. if a leak and/or disconnection occur, ventilation may be adversely affected and the tracheostomy tube might require immediate replacement.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Covidien Shiley Tracheostomy Tube, Cuffed with Inner Cannula
  • Tipo de evento
    Recall
  • ID del evento
    13189
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Covidien NZ Ltd, Level 3, Building 5, Central Park Corporate Centre, 666 Great South Road, Penrose, AUCKLAND 1051
  • Causa
    The manufacturer has received reports of certain size 8 shiley™ reusable cannula, cuffed tracheostomy tubes that have had volume leakage and/or disconnection between the inner and outer cannulae. these events were typically observed during mechanical ventilation. if a leak and/or disconnection occur, ventilation may be adversely affected and the tracheostomy tube might require immediate replacement.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Otto Bock 4R82 Tube Clamp Adapter
  • Tipo de evento
    Recall
  • ID del evento
    13187
  • Fecha de inicio del evento
    2012-07-23
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Ottobock Australia Pty, Suite 1.01 Century Corporate Centre, 62 Norwest Boulevarde, Baulkham Hills, NSW 2153, AUSTRALIA
  • Causa
    The manufacturer received some complaints about adapters that show a crack starting at the clamping of the adapter into direction of the clamping slot. due to the crack, the clamping to the tube adapter is lost and the prosthesis twists when loaded. depending on the situation this may cause the patient to fall.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Philips MRx HeartStart Defibrillator
  • Tipo de evento
    Recall
  • ID del evento
    13185
  • Fecha de inicio del evento
    2012-07-20
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • Causa
    When heartstart mrx defibrillator monitors are used in external transport and ems environments the mechanical/electrical connection between the pads/paddles therapy cable (including pads cpr therapy cable and external paddles cable) and the mrx therapy connection port may experience higher than expected levels of stress causing accelerated wear. without routine periodic inspections and preventive action by users, wear of the connection could ultimately prevent the device from sensing that the pads/paddles therapy cable is connected. this wear also may cause the mrx to inappropriately identify the pads therapy cable, external paddles, or internal paddles.
  • Acción
    Instructions for use to be updated
Retiro De Equipo (Recall) de Roche Modular Pre-Analytics EVO (MPA)
  • Tipo de evento
    Recall
  • ID del evento
    13178
  • Fecha de inicio del evento
    2012-07-19
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Roche Diagnostics (NZ) Ltd, 15 Rakino Way, Mt Wellington, AUCKLAND 1644
  • Causa
    Symantec informed roche in a security advisory about a defect in their product pcanywhere (version 12.6.6 and older), a software product for accessing pcs remotely. the defect is a so called "pre authentication vulnerability" which allows full access to the target system without having to log on.
  • Acción
    Software to be upgraded
Retiro De Equipo (Recall) de Roche Cobas 8000 Data Manager
  • Tipo de evento
    Recall
  • ID del evento
    13178
  • Fecha de inicio del evento
    2012-07-19
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Roche Diagnostics (NZ) Ltd, 15 Rakino Way, Mt Wellington, AUCKLAND 1644
  • Causa
    Symantec informed roche in a security advisory about a defect in their product pcanywhere (version 12.6.6 and older), a software product for accessing pcs remotely. the defect is a so called "pre authentication vulnerability" which allows full access to the target system without having to log on.
  • Acción
    Software to be upgraded
Retiro De Equipo (Recall) de Roche Amplilink
  • Tipo de evento
    Recall
  • ID del evento
    13178
  • Fecha de inicio del evento
    2012-07-19
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Roche Diagnostics (NZ) Ltd, 15 Rakino Way, Mt Wellington, AUCKLAND 1644
  • Causa
    Symantec informed roche in a security advisory about a defect in their product pcanywhere (version 12.6.6 and older), a software product for accessing pcs remotely. the defect is a so called "pre authentication vulnerability" which allows full access to the target system without having to log on.
  • Acción
    Software to be upgraded
Retiro De Equipo (Recall) de Vital Signs Disposable General Purpose Temperature Probe, 9Fr
  • Tipo de evento
    Recall
  • ID del evento
    13150
  • Fecha de inicio del evento
    2012-08-14
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Obex Medical Ltd, 303 Manukau Road, Epsom, AUCKLAND
  • Causa
    The use of the disposable general purpose 9 french temperature probe for nasopharyngeal applications can create, in certain clinical conditions, a potential for severe epistaxis (nose bleed), which may require interventional procedures or prescription medications to properly resolve.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Sysmex CS-2100i/ 2000i Systems
  • Tipo de evento
    Recall
  • ID del evento
    13149
  • Fecha de inicio del evento
    2012-07-21
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, 55 Hugo Johnston Drive, Penrose, Auckland
  • Causa
    Manufacturer advises that cs-2000i and cs-2100i with software version 00-60 show the phenomenon that certain results with error flags highlighted for review, are automatically validated and sent to the host computer without those flags.
  • Acción
    Software to be upgraded
Retiro De Equipo (Recall) de ASP Sterrad 100S System Cassettes
  • Tipo de evento
    Recall
  • ID del evento
    13148
  • Fecha de inicio del evento
    2012-07-11
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
  • Causa
    Recall for a single lot of the sterrad 100s system cassettes. manufacturer has disccovered that printed information on the cassette results in the sterilizer being unable to properly read the barcode. during the automated process used to print information on sterrad 100s system cassettes, an additional thirteenth (13th) bar was added to the barcode resulting in an unusable cassette.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Alcon Acrysof Cachet Phakic Lens
  • Tipo de evento
    Recall
  • ID del evento
    13135
  • Fecha de inicio del evento
    2012-07-10
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Alcon Laboratories Australia Pty Ltd (NZ), c/- Pharmaco (NZ) Ltd, 4 Fisher Crescent, Mt Wellington, Auckland
  • Causa
    Change to instructions for use following review of long term clinical data. changes primarily relate to patient selection and follow-up.
  • Acción
    Instructions for use to be updated
Retiro De Equipo (Recall) de Kimberly-Clark MicroCool Breathable Gown
  • Tipo de evento
    Recall
  • ID del evento
    13134
  • Fecha de inicio del evento
    2012-07-10
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Kimberly-Clark New Zealand, Level 1, 86 Plunket Ave, Manukau City, Auckland
  • Causa
    Outer packaging of individual gowns may be compromised.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Kimberly-Clark Ultra Fabri Reinforced Surgical Gown X-Large
  • Tipo de evento
    Recall
  • ID del evento
    13134
  • Fecha de inicio del evento
    2012-07-10
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Kimberly-Clark New Zealand, Level 1, 86 Plunket Ave, Manukau City, Auckland
  • Causa
    Outer packaging of individual gowns may be compromised.
  • Acción
    Product to be returned to supplier
Retiro De Equipo (Recall) de Radiometer ABL90Flex Blood Gas analysers
  • Tipo de evento
    Recall
  • ID del evento
    13130
  • Fecha de inicio del evento
    2012-07-06
  • País del evento
    New Zealand
  • Fuente del evento
    NZMMDSA
  • URL de la fuente del evento
    https://medsafe.govt.nz/hot/Recalls/RecallSearch.asp
  • Notas / Alertas
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • Notas adicionales en la data
    Recalling Organisation: Radiometer Pacific, Unit 3, 33 Spartan Road, Takanini 2105
  • Causa
    Radiometer medical has become aware that there is a potential risk of reporting of current patient results to a previous patient id. this may occur when an abl90flex analyzer with software versions below v2.7 mr4 is using the hl7 v2.5 communication protocol for transmitting results to a his/lis system.
  • Acción
    Software to be upgraded
  • « First
  • ‹ Prev
  • …
  • 4996
  • 4997
  • 4998
  • 4999
  • Next ›
  • Last »

Acerca de la base de datos

Explore más de 120,000 registros de retiros, alertas y notificaciones de seguridad de dispositivos médicos y sus conexiones con los fabricantes.

  • Preguntas frecuentes
  • Acerca de la base de datos
  • Contáctenos
  • Créditos

Historias en su correo

¿Trabaja en la industria médica? ¿O tiene experiencia con algún dispositivo médico? Nuestra reportería no ha terminado. Queremos oír de usted.

¡Cuéntanos tu historia!

Aviso

Los dispositivos médicos ayudan con el diagnóstico, la prevención y el tratamiento de muchas lesiones y enfermedades. A través de la International Medical Devices Database no estamos sugiriendo que compañías u otras entidades mencionadas en la base de datos hayan sido parte de una conducta ilegal o hayan actuado de manera impropia. Un mismo dispositivo médico puede tener distintos nombres en diferentes países. Esta base de datos no busca proporcionar asesoría médica. Los pacientes deben consultar con sus médicos para determinar si la data contiene información relevante y si la misma tiene implicaciones médicas para ellos.

Descargar la data

La International Medical Devices Database está bajo la licencia Open Database License y sus contenidos bajo la licencia Creative Commons Attribution-ShareAlike . Al usar esta data, siempre citar al International Consortium of Investigative Journalists. Puede descargar acá una copia de la base de datos.

Descargar todo (zipped)