Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Chisel (osteotome) - Product Code KDG
Causa
The stability data does not support the product labeled with a 10-year shelf life.
Acción
The firm initiated their recall by email and letter on 03/30/2017. Consignees were asked to inspect inventory to locate and quarantine affected product. All consignees are asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Mask, oxygen - Product Code BYG
Acción
Recall Letters were sent to consignees on May 10, 2017. The letter requested that they immediately stop distribution and quarantine all affected lots. The letter also requested a sub-recall if the product had been further distributed. Also attached to the letter is a Recall Response Form to be returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Et tube holders contain an adhesive faceplate that may lose its adhesive properties and will not adhere to the patient's face.
Acción
Dale Medical notified accounts (OUS) on March 24, 2017 and US accounts on April 7, 2017 advising them to review inventory, if you have any of the noted Lot numbers in stock contact your customers that may have received the identified lot numbers of 270 ET Tube holders and remove suspected product.. Dale Medical will replace product. For questions customers were instructed to call
508-695-9316 ext. 143
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Catheter,intravascular,therapeutic,short-term less than 30 days - Product Code FOZ
Causa
Arrow international initiated the recall due to reports of safety cover disengagement and needlestick injury for the bd eclipse needle that are packaged with certain arrow products.
Acción
Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Duodenoscope and accessories, flexible/rigid - Product Code FDT
Causa
Fujifilm is conducting a corrective action due to an fda inspection.
Acción
Fujifilm notified their customers on 3/9/2015. Customers are advised of the updated manuals. and asked to destroyed existing manuals. Questions, comments and concerns should be directed to 800.385.4666 between the areas of 8:00 am and 5:00 pm EST or via email at G5info@fujifilm.com.