Causa
Manufacturing defect in the lot number, where several strips with multiple missing pads were found inside the bottle, the use of the strips affected to analyze patient samples can cause faults in the quality control, false negative results for one or several analytes, as well as the ph.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
Loose areas of the test strips in several compartments of the analyzer equipment (such as distributors and strip holders), as well as in the compartments of the strips used. this failure can not be detected by the ichem velocity system and, therefore, can lead to erroneous ph results and false negative results in the following tests: nitrites, glucose, blood, bilirubin, urobilinogen, proteins, lauter-acetic acid, ascorbic acid and ketones.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The reagent does not work as expected on the expiry date. therefore, there is a relationship between the reactive life and the intensity of the coloration, causing false negative or false false results in pathology examinations due to inadequate tissue staining.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer that the level detection system may incorrectly report an "empty bottle" condition during filling, before completing the 5 l filling, leading to potential failure in the results of analysis of the samples and being able to require repeat the biopsies.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer establishes that the internal wiring of the remote alarm plug in the referenced medical devices may be incorrect, which may cause the remote control not to work properly in case of failure of the instrument, the loss of the alarm could lead to improper processing of the shows, which could lead to re-interventions and potential adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer informs that the referenced devices could have an incorrect labeling with respect to the voltage specified for the alarm connector on the back of the equipment, the correct information is 30va.C, however the current labels specify125va.C as maximum voltage, leading to there are electrical risks and possible adverse events on users.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
Potential risk of false negative results when using the reference test kits no. 942-940 for the aqt90flex analyzer and the reference cartridges cal 944-268. the above is due to improper placement of the aluminum films, which affect the absorption capacity of the cartridge component, resulting in premature wear of the product. additionally, this defect can also affect the stability of the reagents contained in the wells of the cartridges, which are susceptible to moisture.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer affirms that medical-referenced devices could report a negative positive result, after the use of an incorrect value of hematocrit, leading to possible errors in diagnosis and adverse events on patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer has detected that the needle may be attached incorrectly to the body of the syringe, in the worst case it could cause the needle to detach, if the needle is separated from the body of the syringe after being removed from a patient, there is risk of an accident with exposure to blood, this situation could lead to possible adverse events on patients and / or users.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
Errors are observed with the bar code when using the batch enmención, because the results of the patient for the test are not delivered, in this sense it is recommended that it be produced using the barcode.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer reports that it has detected in the referenced medical device, changes in the ph content that can cause irritation in the soft tissues of the mouth, leading to potential adverse events in patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer reports that it has detected in the referenced medical device, changes in the ph content that can cause irritation in the soft tissues of the mouth, leading to potential adverse events in patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
The manufacturer reports that he was notified of some cases in which the referenced medical devices do not receive an existing warning or sound alarm, therefore, severe hypoglycemia (low blood sugar level) or severe hyperglycemia (high blood sugar levels) may not be detected. blood), leading to the potential occurrence of adverse events on patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
Interferences of the drugs n-acetylcysteine (nac), acetaminophen and metamizole, administered to patients in therapeutic encounters, in the measurement of analytes based on the so-called trinder reactions. the reactions used a colorimetric reaction between trinder peroxide dehydrogen (h2o2), a phenolic derivative, and amino-antipyrine in the presence of peroxidase, which is catalyzed, reaction interrupted by the drugs mentioned above. dialab states that while this information is updated in the precautions of the insert of the product, these analyzes should be done before the administration of these drugs.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
Assays using the trinder reaction may generate incorrect results in samples containing n-acetylcysteine (nac), paracetamol and metamizole. trinder-based assays using a colorimetric reaction between hydrogen peroxide, a phenol derivative and aminoantipyrine catalyzed in the presence of peroxidase, can be altered by some medications, causing falsely low results.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Notas adicionales en la data
Causa
Introduction of new values for the bio-radusados controls in the liaison cortisol kit and the control systems bio-rad lyphochek® immunoassay plus control, lot40310; bio-rad liquichek ™ urine chemistry control, lot66700 and bio-rad lyphochek® quantitative urine control, lot63350.