Stopcock, i.V. Set - Product Code FMG
Causa
The infusion sets were manufactured with an incorrect spin collar.
Acción
The firm, ICU Medical Inc., sent an "URGENT: Medical Device Recall Notification" letter dated 2/5/2018 to its customers via UPS on 2/5/2018 in the U.S. The letter described the product, problem and actions to be taken. The customers were instructed to: inspect your inventory for specific product and lot numbers; quarantine all affected devices; complete and return the Recall Response Form as applicable (even if you have no affected devices) via Fax to: 1-866-899-7474 or email to: ICUMedical2805@stericycle.com; if you have distributed the product further, notify your accounts that received the product and ask them to contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm ET) to receive a Response Form. Return affected product to Stericycle using labels provided, if you have not received a return label or reply form or require additional assistance, please contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm) the return labels are for single use only. For additional labels visit http://expertezlabel.com.
For further inquiries, please contact ICU Medical Inc. for Global Complaint Management at ProductComplaintsPP@icumed.com or ICU Customer Service at 1-866-829-9025, option 2 (M-F, 8:30am-4pm PT).
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Stopcock, i.V. Set - Product Code FMG
Causa
The infusion sets were manufactured with an incorrect spin collar.
Acción
The firm, ICU Medical Inc., sent an "URGENT: Medical Device Recall Notification" letter dated 2/5/2018 to its customers via UPS on 2/5/2018 in the U.S. The letter described the product, problem and actions to be taken. The customers were instructed to: inspect your inventory for specific product and lot numbers; quarantine all affected devices; complete and return the Recall Response Form as applicable (even if you have no affected devices) via Fax to: 1-866-899-7474 or email to: ICUMedical2805@stericycle.com; if you have distributed the product further, notify your accounts that received the product and ask them to contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm ET) to receive a Response Form. Return affected product to Stericycle using labels provided, if you have not received a return label or reply form or require additional assistance, please contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm) the return labels are for single use only. For additional labels visit http://expertezlabel.com.
For further inquiries, please contact ICU Medical Inc. for Global Complaint Management at ProductComplaintsPP@icumed.com or ICU Customer Service at 1-866-829-9025, option 2 (M-F, 8:30am-4pm PT).
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Stopcock, i.V. Set - Product Code FMG
Causa
The infusion sets were manufactured with an incorrect spin collar.
Acción
The firm, ICU Medical Inc., sent an "URGENT: Medical Device Recall Notification" letter dated 2/5/2018 to its customers via UPS on 2/5/2018 in the U.S. The letter described the product, problem and actions to be taken. The customers were instructed to: inspect your inventory for specific product and lot numbers; quarantine all affected devices; complete and return the Recall Response Form as applicable (even if you have no affected devices) via Fax to: 1-866-899-7474 or email to: ICUMedical2805@stericycle.com; if you have distributed the product further, notify your accounts that received the product and ask them to contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm ET) to receive a Response Form. Return affected product to Stericycle using labels provided, if you have not received a return label or reply form or require additional assistance, please contact Stericycle at 1-866-629-6182 (M-F, 8am-5pm) the return labels are for single use only. For additional labels visit http://expertezlabel.com.
For further inquiries, please contact ICU Medical Inc. for Global Complaint Management at ProductComplaintsPP@icumed.com or ICU Customer Service at 1-866-829-9025, option 2 (M-F, 8:30am-4pm PT).
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. the collimator may be partially supported by the cart (e.G. leaning against the cart), or it may fall/slide all the way to the floor.
Acción
Philips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
The ldpe bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the ldpe bag on the implant after it is removed from the bag.
Acción
A similar recall was initiated in January 2016 to remove remaining inventory packaged in the old bag. Further evaluation identified additional lots packaged in the old bag that were not included in the January 2016 recall; therefore, resulting in this new recall for the additional lots packaged in the old bag.
A firm, Zimmer Biomet, sent an "Urgent Medical Device Recall" notices dated November 29, 2017 and response forms to customers. The notice described the product, problem and actions to be taken. The Customers should take the following action:
Risk Manager Responsibilities:
1. Review this notification and ensure that affected personnel are aware of the contents.
2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.
3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products at your facility.
4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation.
Distributor Responsibilities:
1. Review this notification and ensure that affected team members are aware of the contents.
2. Immediately locate and quarantine affected product in your inventory.
3. Immediately return all affected product from your distributorship and from affected hospitals within your territory.
a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days.
b. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com.
c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing.
d. Mark RECALL on the outside of the returned cartons.
4. Return the Additi
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. the collimator may be partially supported by the cart (e.G. leaning against the cart), or it may fall/slide all the way to the floor.
Acción
Philips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. the collimator may be partially supported by the cart (e.G. leaning against the cart), or it may fall/slide all the way to the floor.
Acción
Philips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. the collimator may be partially supported by the cart (e.G. leaning against the cart), or it may fall/slide all the way to the floor.
Acción
Philips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
An eeprom chip error code may occur when the catheters are used with carto ep navigation systems.
Acción
Stryker sent an Urgent Medical Device Recall letter dated January 2, 2018, explains the reason for recall, risk to health, and instructs them to discontinue use of the affected products. The customer is to check their inventory for the affected serial numbers provided in the attachment and complete the recall effectiveness check form to indicate the amount of product in inventory. Return instructions are provided. For further questions, please call (888) 763-8803.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Affected lot of 5 ml oral dispensers was packaged in individual pouches that are mislabeled with the incorrect product code (h93876101) and volume (1ml).
Acción
Customers were notified via letter on about 02/06/2018. Instructions include to locate and remove affected product in inventory, confirm mislabeled 5 mL oral dispensers (Lot number 1203395) in the individual pouch are not mixed together with 1 mL oral dispensers (Product code H93876101) or other stored product, contact Baxter Healthcare Center to arrange return of affected product, and complete and return the customer reply form if purchased directly from Baxter. Also, customer are instructed to notify customers and conduct a consumer-level recall if the affected product was further distributed.
Questions and concerns can be directed to the Public Contact, Center for One Baxter with phone number (Toll Free) 800-422-9837 and 847-948-4770, Monday through Friday 8:00 AM to 5:00 PM CST.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Interventional fluoroscopic x-ray system - Product Code OWB
Causa
Extra monitors, so called 7th or 8th monitor, mounted on the monitor ceiling suspension (mcs), might disengage from the mcs and fall down. the two small set screws securing the two gliding rods into an aluminum block of
the support arm can come loose because of vibrations and forces exerted on them. if the set screws have come loose gravity can keep the rods in position. however, if a force is exerted on the monitor pushing it forward, the monitor can
topple forward. once the monitor has toppled over it is only attached with its wires. if the monitor breaks loose of its wires it will drop on the floor.
Acción
Philips sent an Urgent Field Safety Notice letter December 20, 2017 dated to all affected customers. The letter identified the affected product, problem and the actions to be taken. For question contact your local Philips representative.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
An eeprom chip error code may occur when the catheters are used with carto ep navigation systems.
Acción
Stryker sent an Urgent Medical Device Recall letter dated January 2, 2018, explains the reason for recall, risk to health, and instructs them to discontinue use of the affected products. The customer is to check their inventory for the affected serial numbers provided in the attachment and complete the recall effectiveness check form to indicate the amount of product in inventory. Return instructions are provided. For further questions, please call (888) 763-8803.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
The affected agc 2000 lp patella products were over-molded leaving the outside diameter nonconforming to the print dimension.
Acción
Consignees were notified via mail on approximately 08/25/2017. Instructions to distributors, sales representatives, and distributor operations managers include to ensure affected team members are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide any additional hospitals and/or surgeons that received direct shipments.
Instructions to risk managers and surgeons include to ensure affected personnel are informed of the recall, quarantine all affected product, return any affected product in inventory, and complete and return the Certificate of Acknowledgment.
Questions or concerns can be sent to the Zimmer Biomet customer call center at 1-574-371-3071 between 8:00 am and 5:00 pm EST, Monday through Friday. Or they can be sent via email to corporatequality.postmarket@zimmerbiomet.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
The affected agc 2000 lp patella products were over-molded leaving the outside diameter nonconforming to the print dimension.
Acción
Consignees were notified via mail on approximately 08/25/2017. Instructions to distributors, sales representatives, and distributor operations managers include to ensure affected team members are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide any additional hospitals and/or surgeons that received direct shipments.
Instructions to risk managers and surgeons include to ensure affected personnel are informed of the recall, quarantine all affected product, return any affected product in inventory, and complete and return the Certificate of Acknowledgment.
Questions or concerns can be sent to the Zimmer Biomet customer call center at 1-574-371-3071 between 8:00 am and 5:00 pm EST, Monday through Friday. Or they can be sent via email to corporatequality.postmarket@zimmerbiomet.com.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Interventional fluoroscopic x-ray system - Product Code OWB
Causa
Potential for liquid entry.
Acción
On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail.
Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Interventional fluoroscopic x-ray system - Product Code OWB
Causa
There is an increase in the failure rate of certain anode drive units (adu5) used in these products. early life failure of the adus only occurs when the hospital mains provide 480v to the system. this current voltage may lead to saturation and overheating of the coils that protect the igbts (insulated-gate bipolar transistor) of the adu. the overheating may also generate a peculiar burning smell that may be noticed. in some instances this burning smell may be noticed before the adu fails. when the adu fails the x-ray performance of the system is reduced to
emergency fluoroscopy. exposure will not be possible and image quality is reduced.
Acción
Customer Notification letter dated August 2017 were sent to affected customers. The letter identified affected product, problem and actions to be taken. The customer is requested to place this Field Safety Notice with the documentation of the system until this correction has been implemented. For questions contact your local Philips representative.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Wrap, sterilization - Product Code FRG
Causa
On june 15, 2017, steris identified that the purple ink located on the outside label of vis-u-all pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Acción
Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.