• Acerca de la base de datos
  • ¿Cómo usar la IMDD?
  • Descargar la data
  • Preguntas frecuentes
  • Créditos
Vista de la lista Vista de las tarjetas
  • Dispositivo 960
  • Fabricante 297
  • Evento 124969
  • Implante 79
Retiro De Equipo (Recall) de Integra LifeSciences
  • Tipo de evento
    Recall
  • ID del evento
    60300
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0357-2012
  • Fecha de inicio del evento
    2011-10-14
  • Fecha de publicación del evento
    2011-12-02
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-03-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105222
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Light source, fiberoptic, routine - Product Code FCW
  • Causa
    A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
  • Acción
    Integra sent an URGENT: Field Corrective Action letter dated October 14, 2011 to all affected customers via FedEx. The letter identified the affected product, the problem, and the actions to be taken with instructions to immediately inspect their inventory for potentially loose screws within the affected unit(s). If the unit(s) were observed with loose screws, the unit(s) were to be returned to Integra Surgical for repair and servicing. Customers were instructed to complete the Acknowledgement and Return Form per the instructions on the form. For questions call Integra Surgical Customer Service at 1-800-431-1123.
Retiro De Equipo (Recall) de Integra LifeSciences
  • Tipo de evento
    Recall
  • ID del evento
    60300
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0358-2012
  • Fecha de inicio del evento
    2011-10-14
  • Fecha de publicación del evento
    2011-12-02
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-03-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105223
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Light source, fiberoptic, routine - Product Code FCW
  • Causa
    A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
  • Acción
    Integra sent an URGENT: Field Corrective Action letter dated October 14, 2011 to all affected customers via FedEx. The letter identified the affected product, the problem, and the actions to be taken with instructions to immediately inspect their inventory for potentially loose screws within the affected unit(s). If the unit(s) were observed with loose screws, the unit(s) were to be returned to Integra Surgical for repair and servicing. Customers were instructed to complete the Acknowledgement and Return Form per the instructions on the form. For questions call Integra Surgical Customer Service at 1-800-431-1123.
Retiro De Equipo (Recall) de Genetic Screening Processor
  • Tipo de evento
    Recall
  • ID del evento
    60390
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0603-2012
  • Fecha de inicio del evento
    2011-10-24
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-07-20
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105229
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Radioimmunoassay, thyroid-stimulating hormone - Product Code JLW
  • Causa
    Malfunction of the gsp disk remover unit may cause a punched disc to remain in the well during the measurement of the plate. this may cause a false negative screening result being reported for 17-ohp, resulting in failure to identify cases of congenital adrenal hyperplasia (cah).
  • Acción
    Recall (correction) letters were sent to all customers on October 24, 2011. An improved disk remover design is under development to eliminate the reported problem. Until the final field correction is available, customers should follow the precautions provided. The needed protocol update will be performed by the local service engineer. If customers observe an increase in the occurrence of the problem, the laboratory should contact their local PerkinElmer representative.
Retiro De Equipo (Recall) de RayStation
  • Tipo de evento
    Recall
  • ID del evento
    60409
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0359-2012
  • Fecha de inicio del evento
    2011-10-19
  • Fecha de publicación del evento
    2011-12-02
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-05-11
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105264
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System,planning,radiation therapy treatment - Product Code MUJ
  • Causa
    The firm has found an issue with dicom rt image representation of drr images from raystation 2.0 which needs to be clarified in the user manual. the symptoms involve the pixel spacing and image position attributes in the rt images being incorrect, thus enabling the user to make measurements that are erroneous and possibly misalign the patient.
  • Acción
    The firm, RaySearch Laboratories, sent an email with an "Urgent Field Safety Notice Medical Device Correction" notice dated October 19, 2011 to its customers. The notice describes the product, problem and actions to be taken. The customers were instructed to pass on the information in the field safety notice to everyone involved and create routines to compensate for the incorrect DRR images; if the DICOM DRR images are not used to make any form of absolute measurements, no action needs to be taken. Raysearch laboratories will provide a script to update these DICOM files to have the correct values upon request. For further information please contact RaySearch Laboratories AB Quality and Regulatory Affairs Manager at +46 722 366 110 or david.hedfors@raysearchlabs.com; RaySearch Americas Director of Customer Support at +1 877 778 3489 or freddie.cardell@raysearchlabs.com.
Retiro De Equipo (Recall) de TrueBeam and TrueBeam STx,
  • Tipo de evento
    Recall
  • ID del evento
    60411
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0364-2012
  • Fecha de inicio del evento
    2011-10-31
  • Fecha de publicación del evento
    2011-12-02
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-07-25
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105265
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel data.
  • Notas adicionales en la data
    Accelerator, linear, medical - Product Code -IYE
  • Causa
    An anomaly has been identified with the truebeam and truebeam stx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.
  • Acción
    On 11/04/2011, an Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps. Varian is developing a corrective action for this issue. A customer service representative will contact you when the correction is available to schedule installation on your system. Questions regarding the letter should be directed to the Varian Oncology Help Desk Contact Information at USA and Canada: 1-888-827- 4265 or Europe +41-41-749-8844.
Retiro De Equipo (Recall) de McKesson MediPak
  • Tipo de evento
    Recall
  • ID del evento
    60412
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0490-2012
  • Fecha de inicio del evento
    2011-09-02
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-10-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105298
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter, urological - Product Code KOD
  • Causa
    Nurse assist, inc. received a recall notice from the triad group. this recall has been initiated due to a recall of the pvp solution packs manufactured by the triad group that are contained in certain procedure trays.
  • Acción
    Nurse Assist sent a recall notice dated September 2, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory immediately and quarantine the affected product. Customers were also instructed to identify their customers and notify them at once of the product recall. Customers were instructed to complete and return the enclosed response form as soon as possible. For those distributors that choose to destroy the affected product they must use the attached Notice of Destruction to report the specifics of the destroyed product. and contact Nurse Assist Customer Service for a RGA # prior to destruction. For any questions call Nurse Assist Customer Service at 1-800-649-6800.
Retiro De Equipo (Recall) de SoftReports
  • Tipo de evento
    Recall
  • ID del evento
    60416
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0362-2012
  • Fecha de inicio del evento
    2011-10-25
  • Fecha de publicación del evento
    2011-12-02
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-03-16
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105363
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Medical computers and software - Product Code LNX
  • Causa
    On 10/25/2011 scc soft computer, clearwater, fl initiated a correction on the softreports version 1.1.8.2.4 with hot fixes 1.7199.7 or 1.7200.7. when softreports designer opened a layout, the system did not load the layout with the previously saved value in the newpage property. this caused multiple patient records to be printed on the same page. all affected clients were notified on the issue on.
  • Acción
    SCC Soft Computer sent a "Correction Notification" letter dated October 25, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The hot fixes that were previously loaded which introduced this issue were corrected and delivered to all affected customers as a mandatory correction in hot fixes 1.7199.8 and 1.7200.8 and higher.
Retiro De Equipo (Recall) de McKesson MediPak
  • Tipo de evento
    Recall
  • ID del evento
    60412
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0491-2012
  • Fecha de inicio del evento
    2011-09-02
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-10-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105373
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter, urological - Product Code KOD
  • Causa
    Nurse assist, inc. received a recall notice from the triad group. this recall has been initiated due to a recall of the pvp solution packs manufactured by the triad group that are contained in certain procedure trays.
  • Acción
    Nurse Assist sent a recall notice dated September 2, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory immediately and quarantine the affected product. Customers were also instructed to identify their customers and notify them at once of the product recall. Customers were instructed to complete and return the enclosed response form as soon as possible. For those distributors that choose to destroy the affected product they must use the attached Notice of Destruction to report the specifics of the destroyed product. and contact Nurse Assist Customer Service for a RGA # prior to destruction. For any questions call Nurse Assist Customer Service at 1-800-649-6800.
Retiro De Equipo (Recall) de Siemens MAMMOMAT Inspiration
  • Tipo de evento
    Recall
  • ID del evento
    60423
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0523-2012
  • Fecha de inicio del evento
    2011-09-15
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2014-03-05
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105375
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Full field digital,system,x-ray,mammographic - Product Code MUE
  • Causa
    Lowering the swivel arm with "object table" while an operator's or patient's body parts are underneath the "object table" may cause the "object table" to collide with and injure the operator or patient. therefore attention should be paid when adjusting "object table" height to avoid a possible risk of injury to the operator or the patient. the movement of the swivel arm may also cause an injur.
  • Acción
    Siemens Medical Solutions USA, Inc sent a Safety Advisory Notice dated January 8, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The notification included an Addendum to the User Manual. We appreciate your understanding and cooperation with this safety advisory and ask you to immediately instruct your operating personnel accordingly. Please ensure that this safety advisory is placed in the System's Operator Manual. Your operating personnel should maintain awareness over an appropriate defined period. Further questions please call (610) 219-4834.
Retiro De Equipo (Recall) de Siemens syngo Imaging XS, versions VA70A and VA70B Client/Server Ins...
  • Tipo de evento
    Recall
  • ID del evento
    60424
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0515-2012
  • Fecha de inicio del evento
    2011-09-19
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2014-01-24
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105377
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, image processing, radiological - Product Code LLZ
  • Causa
    Unintended behavior when using syngo imaging xs, versions va70a and va70b client/serve installations. loading additional images/series from patient browser using the thumbnails can cause an unintended deletion of images/series.
  • Acción
    Siemens Medical Solutions USA, Inc sent a Customer Safety Advisory letter dated August 24, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you detect patient data inconsistenceis in your system caused by this problem, please contact Siemens Service for support. For further questions please call (610) 219-4834.
Retiro De Equipo (Recall) de Siemens syngo Imaging XS, all versions
  • Tipo de evento
    Recall
  • ID del evento
    60425
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0514-2012
  • Fecha de inicio del evento
    2011-09-20
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2014-03-05
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105378
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, image processing, radiological - Product Code LLZ
  • Causa
    Unintended behavior when using syngo imaging xs, all versions. the "rename" functionality can cause an unintended patient merge if the renamed patient is sent to another system.
  • Acción
    Siemens Medical Solutions USA, Inc sent a Customer Safety Advisory Notice letter dated August 3, 2011 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If you detect patient data inconsistencies in your system caused by this problem, please contact Siemens Service for support. For further questions please call (610) 219-4834.
Retiro De Equipo (Recall) de Siemens syngo.plaza
  • Tipo de evento
    Recall
  • ID del evento
    60426
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0522-2012
  • Fecha de inicio del evento
    2011-09-21
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2014-01-31
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105379
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    system image processing radiological - Product Code LLZ
  • Causa
    Potential malfunction when using syngo.Plaza. on syngo.Plaza workplaces large studies can be loaded into the viewer if transferred from a dicom node or imported even while image transfer is not yet completed. in these situations more images of that study may exist which are not shown in the syngo.Plaza viewer.
  • Acción
    Siemens Medical Solutions USA sent a Customer Safety Advisory Notice dated March 10, 2011 to the affected customer. The letter identified the product the problem and the action needed to be taken by the customer. The letter informed the customer of the potential issues and provides additional information on the implemented resolution. For further questions please call (610) 219-4834.
Retiro De Equipo (Recall) de Truform Therapeutic Classic Medical Style Compression Pantyhose
  • Tipo de evento
    Recall
  • ID del evento
    60371
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0392-2012
  • Fecha de inicio del evento
    2011-10-07
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-03-05
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105382
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Stocking, medical support (for general medical purposes) - Product Code FQL
  • Causa
    During an audit with the fda, it was discovered that a new compression testing device used for testing hosiery had been implemented and that the operators were not properly trained. additionally, the acceptance criteria for the new device were not changed from the old testing device to reflect additional data (compression profile percentages). this resulted in the device history files containing s.
  • Acción
    The firm, Surgical Appliance Industries sent an URGENT: MEDICAL DEVICE RECALL letters dated November 7, 2011 to all affected customers. The letter described the products, problem and actions to be taken. Customers were instructed to immediately examine their inventory and quarantine all product(s) subject to recall for return and exchange. In addition, the letter states that if the product was further distributed to contact their customers and notify them of the recall. Customers were asked to complete and return the enclosed Recall Response Form. For any questions contact Quality Systems Manager, at 513-271-4594 ext. 282.
Retiro De Equipo (Recall) de Stryker EMS Stair PRO
  • Tipo de evento
    Recall
  • ID del evento
    60422
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0943-2012
  • Fecha de inicio del evento
    2011-10-31
  • Fecha de publicación del evento
    2012-02-06
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-08-28
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105384
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Stretcher, hand-carried - Product Code FPP
  • Causa
    The foot rest kit may have the wrong pivot spacers making the joint more susceptible to relaxation and prone to fatigue. there is a moderate risk of injury to the patient should the handles become detached during use.
  • Acción
    Stryker Medical spoke with all customers who received these shipments prior to 10/31/2011 requesting information on kits affected and recalled foot rest units installed. December 5, 2011 Urgent Medical Device Recall Letters, with instructions on how to to either install a new Foot Rest Kit ( enclosed ) or dispose of any kits that were not installed were sent to both customers who could have installed recalled foot rest units. Customers needing help or assistance with installation are asked to call Stryker Technical Support a 1-800-327-0770. Customers were to complete and return a postage paid postcard confirming receipt. Customers with questions or concerns are asked to contact Kristin Dudek at 269-389-6927 M-F m 8 a.m. - 5 p.m. (EST)
Retiro De Equipo (Recall) de WOLFPAK REDIFLO ELASTOMERIC INFUSION PUMP
  • Tipo de evento
    Recall
  • ID del evento
    60316
  • Clase de Riesgo del Evento
    Class 1
  • Número del evento
    Z-0269-2012
  • Fecha de inicio del evento
    2011-10-20
  • Fecha de publicación del evento
    2011-11-23
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-04-25
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105401
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    ELASTOMERIC INFUSION PUMP - Product Code MEB
  • Causa
    Redi-flo elastomeric infusion pumps may have a higher flow rate than specified (more than +15%).
  • Acción
    On 10/20/2011 Wolf Medical Supply Inc., customers were notified by telephone and email. Product was distributed to end users.
Retiro De Equipo (Recall) de Eclipse Treatment Planning System
  • Tipo de evento
    Recall
  • ID del evento
    60441
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0370-2012
  • Fecha de inicio del evento
    2011-10-31
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2014-06-06
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105415
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    System, planning, radiation therapy treatment - Product Code MUJ
  • Causa
    An anomaly has been identified with varian eclipse and rt chart where adding a reference point in rt chart and editing the field dose contributions may change the prescribed dose per fraction in eclipse.
  • Acción
    Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION/URGENT FIELD SAFETY NOTICE" dated October 31, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Varian has corrected the issue and will contact customers to schedule installation of the corrected version of the system. Contact the Varian Oncology Help Desk at 1-888-827-4265 for questions regarding this notice. .
Retiro De Equipo (Recall) de Anthem CRTP
  • Tipo de evento
    Recall
  • ID del evento
    60134
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0430-2012
  • Fecha de inicio del evento
    2011-09-23
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-04-10
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105436
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Implantable pacemaker pulse-generator - Product Code DXY
  • Causa
    St. jude medical has received reports of out of range pacing lead impedance (pli) measurements. these measurements are low and out of range (<200 ohms) but measure in range during visits to the clinic. the out of range readings are concluded as false.
  • Acción
    The firm, St. Jude Medical, sent an "IMPORTANT PRODUCT FIELD CORRECTION" letter dated September 22, 2011 to its customers by FedEx priority overnight service for planned receipt on September 23, 2011. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed that in order to prevent a false reading, a new programmer software version (12.1.0.1) is available. This software when used to interrogate an Accent DR or Anthem CRT-P pacemaker will eliminate the potential for this anomaly to occur. The onetime upgrade is performed automatically on affected devices and will not change the operation of the implanted device. A St. Jude Medical Sales Representative will assist the customers in loading the new programmer software onto their Merlin programmer. Customers with questions or concerns, can contact their local St. Jude Medical representative or their Technical services department at (800) 722-3774.
Retiro De Equipo (Recall) de Urine Reagent Strips (URS) 10 Parameters
  • Tipo de evento
    Recall
  • ID del evento
    60476
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0511-2012
  • Fecha de inicio del evento
    2011-10-14
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-04-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105522
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Automated urinalysis system - Product Code KQO
  • Causa
    This recall was initiated in response to findings during a fda inspection of the firm. the firm's recall has been initiated due to improper manufacture of sub-component nitrite paper lot 35134, expiration date: 01/2014. discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets (refer to next paragraph). sampling size submitted for qc testing was insuff.
  • Acción
    Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.
Retiro De Equipo (Recall) de Urine Reagent Strips (URS) 10 Parameters
  • Tipo de evento
    Recall
  • ID del evento
    60476
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0512-2012
  • Fecha de inicio del evento
    2011-10-14
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-04-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105523
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Automated urinalysis system - Product Code KQO
  • Causa
    This recall was initiated in response to findings during a fda inspection of the firm. the firm's recall has been initiated due to improper manufacture of sub-component nitrite paper lot 35134, expiration date: 01/2014. discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets. sampling size submitted for qc testing was insufficient based upon number o.
  • Acción
    Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.
Retiro De Equipo (Recall) de Synergy XVI kV Generator
  • Tipo de evento
    Recall
  • ID del evento
    60497
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0342-2012
  • Fecha de inicio del evento
    2011-10-20
  • Fecha de publicación del evento
    2011-12-01
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-08-09
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105583
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Accelerator, linear, medical - Product Code IYE
  • Causa
    A small number of sites have experienced situations with kv generator transformers overheating due to a diode failure on the interface board.
  • Acción
    The firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.
Retiro De Equipo (Recall) de Biomet Series A Thin Patella 25mm, three pegs, Part 184780
  • Tipo de evento
    Recall
  • ID del evento
    60144
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0695-2012
  • Fecha de inicio del evento
    2011-10-04
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-04-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105815
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prothesis, knee,patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Correcting language on product labeling which incorrectly states patella buttons are intended for use with the agc knee system.
  • Acción
    Biomet sent a "FIELD CORRECTION NOTICE" dated October 3, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected product and to follow the work instructions included in the letter. Additionally, a Confirmation of Completion Response Form was enclosed for customers to complete and return via fax to 574-372-1683. Questions related to this notice are directed to 574-371-3755.
Retiro De Equipo (Recall) de Biomet Series A Standard Patella size 25mm, one peg 184700
  • Tipo de evento
    Recall
  • ID del evento
    60144
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0696-2012
  • Fecha de inicio del evento
    2011-10-04
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-04-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105817
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prothesis, knee,patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Correcting language on product labeling which incorrectly states patella buttons are intended for use with the agc knee system.
  • Acción
    Biomet sent a "FIELD CORRECTION NOTICE" dated October 3, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected product and to follow the work instructions included in the letter. Additionally, a Confirmation of Completion Response Form was enclosed for customers to complete and return via fax to 574-372-1683. Questions related to this notice are directed to 574-371-3755.
Retiro De Equipo (Recall) de Kendall
  • Tipo de evento
    Recall
  • ID del evento
    30196
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0146-05
  • Fecha de inicio del evento
    2004-10-08
  • Fecha de publicación del evento
    2004-11-02
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2012-05-15
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=35499
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Needle, Hypodermic, Single Lumen - Product Code FMI
  • Causa
    Needle cannula may separate from the hub during use.
  • Acción
    Recall letters requesting product return were mailed to all customers on October 8, 2004
Retiro De Equipo (Recall) de Biomet Series A Standard Patella size 28mm, one peg 184702
  • Tipo de evento
    Recall
  • ID del evento
    60144
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0697-2012
  • Fecha de inicio del evento
    2011-10-04
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-04-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105818
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prothesis, knee,patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Correcting language on product labeling which incorrectly states patella buttons are intended for use with the agc knee system.
  • Acción
    Biomet sent a "FIELD CORRECTION NOTICE" dated October 3, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected product and to follow the work instructions included in the letter. Additionally, a Confirmation of Completion Response Form was enclosed for customers to complete and return via fax to 574-372-1683. Questions related to this notice are directed to 574-371-3755.
Retiro De Equipo (Recall) de Biomet Series A Thin Patella size 25mm, one peg 184720
  • Tipo de evento
    Recall
  • ID del evento
    60144
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0698-2012
  • Fecha de inicio del evento
    2011-10-04
  • Fecha de publicación del evento
    2012-01-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2013-04-08
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=105823
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prothesis, knee,patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Correcting language on product labeling which incorrectly states patella buttons are intended for use with the agc knee system.
  • Acción
    Biomet sent a "FIELD CORRECTION NOTICE" dated October 3, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to locate the affected product and to follow the work instructions included in the letter. Additionally, a Confirmation of Completion Response Form was enclosed for customers to complete and return via fax to 574-372-1683. Questions related to this notice are directed to 574-371-3755.
  • « First
  • ‹ Prev
  • …
  • 4
  • 5
  • 6
  • 7
  • 8
  • …
  • Next ›
  • Last »

Acerca de la base de datos

Explore más de 120,000 registros de retiros, alertas y notificaciones de seguridad de dispositivos médicos y sus conexiones con los fabricantes.

  • Preguntas frecuentes
  • Acerca de la base de datos
  • Contáctenos
  • Créditos

Historias en su correo

¿Trabaja en la industria médica? ¿O tiene experiencia con algún dispositivo médico? Nuestra reportería no ha terminado. Queremos oír de usted.

¡Cuéntanos tu historia!

Aviso

Los dispositivos médicos ayudan con el diagnóstico, la prevención y el tratamiento de muchas lesiones y enfermedades. A través de la International Medical Devices Database no estamos sugiriendo que compañías u otras entidades mencionadas en la base de datos hayan sido parte de una conducta ilegal o hayan actuado de manera impropia. Un mismo dispositivo médico puede tener distintos nombres en diferentes países. Esta base de datos no busca proporcionar asesoría médica. Los pacientes deben consultar con sus médicos para determinar si la data contiene información relevante y si la misma tiene implicaciones médicas para ellos.

Descargar la data

La International Medical Devices Database está bajo la licencia Open Database License y sus contenidos bajo la licencia Creative Commons Attribution-ShareAlike . Al usar esta data, siempre citar al International Consortium of Investigative Journalists. Puede descargar acá una copia de la base de datos.

Descargar todo (zipped)