Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
2 customer complaints in total were received from distributors located in germany and rance with regard to inconsistent results from the expected results on the destiny plus analyzer whilst running a prothrombin time (pt assay). these incidences were noted on the mechanical mode of the instrument.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
One of the tyvek breather pouches supplied to medline by alcan packaging may have an inadequate seal that may cause a compromise in the sterility of the product.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The manufacturer's supplier incorrectly labeled product cartons with a 5-year expiry date instead of a 3-year date.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Low positive control range for the low positive control (5760-4-003078) included in the elisa kit is incorrect. the correct range for the control is (0.91-2.3).
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
One of the components of the iviewgt retractable arm midle arm assembly may be subject to failure overtime. at a certain state the failure will cause the midle arm to drop and be hazardous for patient.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
There is a possibility for lifecycle risk calculation errors to occur when using specific software versions.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Mdl was cancelled on march 12 2004 because of an invalid qa certificate.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Msup (battery) exchange to enhance performance and unit life.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Depuy has identified the potential for the sigma hp fixed bearing tibial tray impactor to become unintentionally locked on the tibial tray implant when impacting the tray into position during surgery due to the interference of the protective cover on the tray.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Possibilite que la sterilite soit compromise en raison d'une dechirure minime au niveau du sceau de l'enveloppe externe.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Electronics box on some flexicoils can overheat and may cause a skin burn when in direct contact with the patient.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Manufacturer uncovered a defect in a small percentage of bag decanters as the decanters were falling apart during use.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
In march 2017 as part of field action 2017-001-c pentax medical informed customers about a potential cracks and gaps in the distal cap of the ed-3490tk and instituted a distal cap inspection program. during discussion with the us fda as part of this field action pentax medical agreed to redesign the ed-3490tk's forceps elevator mechanism o-ring and distal end to reduce the risk of cross-contamination resulting from body fluids entering portions of the ed-3490tk that cannot be easily cleaned.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Cell 15 of panocell-16 lot 13096 is demonstrating weak positive reactions in direct antiglobulin tests.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Product performance is inconsistent with other clinical findings. unacceptable precision patient values may be significantly elevated or depressed.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
If skip forward button used during stage 2 segmentation - system may calculate final dose with the pencil beam algorithm when the monte carlo algorithm is selected with no indication of this occurring.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
A certain error message is displayed when a transducer is disconnected reconnected or another transducer is selected. if this error message is displayed and the system continues to be used the surface of the acoustic lens of the transducer may become hot which can result in mild burns if the transducer is used on one area of the body for a long time.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Conmed linvatec is voluntarily recalling selected serial numbers due to the front panel of the console containing incorrect artwork depicting a bf rating (man in box) per iec60601-1: medical electrical equipment: general requirements for basic safety and essential performance.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Volcano corporation is initiating a voluntary field corrective action because an issue has been identified with sofware versions v3.2.X v3.3 and v3.4 that runs on the s5/s5i/core/core mobile systems ("impacted systems"). this voluntary recall only affects impacted systems. volcano corporation has become aware of a system incompatibility issue between impacted systems and hospital network scans.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
The reason for the recall is related to revised hemolysis claims for the alanine aminotransferase both with and without pyridoxal activation.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
There have been reports of sporadic image errors during patient examination. the image is displayed as completely black. the image quality is not sufficient for diagnosis and the examination would be repeated.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
Recall to upgrade software version to reduce the maximum plasma pulse rate and add a new display that counts pulse used per patient treatment. complaints were received which were attributed to over treatment.
Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
Notas adicionales en la data
Causa
During the investigation of an issue from a customer complaint it was found by the manufacturer that the daily cleanup sample can fail in deleting samples which are older than the configured retention period in the cobas it middleware. some samples which have been ordered from the lis and remain with open tests may not be deleted. open tests are those for which results did not complete the processing cycle including transmission to lis.