Recalling Organisation: Beckman Coulter NZ Ltd, Suite 2, Level 4, 30 St Benedicts Street, Newton, AUCKLAND
Causa
An internal investigation indicates that specific lots of check valves used in several locations with the lh750 and lh780 analysers may fail. this failure may result in an air or liquid leak at the vcs waste chamber or probe waste chamber.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Johnson & Johnson Medical New Zealand Ltd, 507 Mt Wellington Highway, Mt Wellington, AUCKLAND 1060
Causa
The manufacturer has confirmed a small number of tab failures and fascial dehiscences in lower abdominal incisions involving this product.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Elite Scooters (NZ) Ltd, 79 Fairview Road, Katikati, Bay Of Plenty 3129
Causa
The breeze c scooter may become unresponsive during operation due to an electrical contact which may fault due to extensive wear, which potentially could result in the driver not being able to stop the scooter during operation except by turning off the ignition key.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Medtronic New Zealand Limited, Unit 16 Mezzanine Level, 5 Gloucester Park Road, Onehunga, AUCKLAND
Causa
A medtronic investigation has determined that a subset of 2090 programmers have an incorrect software configuration for their region. all programmers contain software that has been validated and released by medtronic; however, affected programmers may contain additional programming options that are not enabled for implanted devices in the specific region.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Causa
Ge healthcare has recently become aware of a potential safety issue due to ecg filter settings associated with the b20/b40 v1, b30 and b40 v2 patient monitors. the b20, b30 and b40 patient monitors set the ecg filter setting incorrectly if the monitor starts with impedance respiration set in the on position. instead of the user selected filter, an additional 0.5 hz high pass filter is used. the additional 0.5 hz high pass filter causes changes to the ecg waveform morphology and incorrect display of st values. the st-elevation measurement can be under estimated in this situation and therefore the treatment of the patient can be delayed.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
It has been determined that specific shipping configurations in which more than eight (8) 6 ounce units which are shipped together with a specific revision of the instructions for use (ifu) can place pressure on the package, which can lead to damage and leakage.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
During a rad examination with automatic exposure control, a highly unlikely malfunction can result in radiation continung longer than required.This may also result in an overexposed image that is not of diagnostic quality, so that th examination needs to be repeated.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
During a rad examination with automatic exposure control, a highly unlikely malfunction can result in radiation continung longer than required.This may also result in an overexposed image that is not of diagnostic quality, so that th examination needs to be repeated.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
During a rad examination with automatic exposure control, a highly unlikely malfunction can result in radiation continung longer than required.This may also result in an overexposed image that is not of diagnostic quality, so that th examination needs to be repeated.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Causa
While operating in presure control ventilation - volume guarantee (pcv-vg) mode, the affected products may produce an over delivery tidal volume when given a unique sequence of user inputs and a collapsed bellows (partially filled circuit).Exposure to excessive tidal volumes while pressure limits are maintained can result in pulmonary volutrauma and/or an increased pulmonary vascular resistance that can negatively affect cardiac filling pressure and/or stroke volume.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: InterMed Medical Ltd, 71 Apollo Drive, Albany, AUCKLAND 1311
Causa
Analysis from customer complaints has revealed that the manufacturer-defined limits for the loudspeaker in the unit were exceeded. consequently a defective loudspeaker will result in an alarm message and alarm buzzer.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
Siemens healthcare diagnostics has confirmed that under a rare set of conditions when utilizing the routine inventory screen to enter a user defined method (empty) flex the system may assign the user defined method flex to a different flex that is currently in inventory on the system, and then use the incorrect flex cartridge to process the user defined method. investigation of the issue has determined that this error will only occur if the assign empty reagent cartridge window has been left open and the flex inventory on the system changes.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Carestream Health New Zealand Ltd, c/- Quigg Partners, Level 7, 28 Brandon St, Wellington
Causa
Carestream health has received one report from overseas where under very specific conditions, some images from multi-slice studies such as ct and mr may not load into the viewer application. an incorrect number of images loaded may potentially lead to an incomplete study report from those images in v11.3.2.X and v11.4 installations in circumstances where usage of multi-slice data is part of the workflow scenarios. this problem occurs infrequently and only when all of the following conditions are met:, 1) the version installed is v11.3.2 or v11.3.2.4 or v11.4, 2) the exam type is from multi-slice modalities, 3) the images are requested to load into a pre-defined hanging protocol where at least one of the hanging protocol groups includes more than 100 images., 4) a timing conflict occurs between 2 processes responsible for the load of images into memory.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent. hospira requests that users do not use incorrectly sized receptal liners with receptal canisters.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent. hospira requests that users do not use incorrectly sized receptal liners with receptal canisters.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: BioMerieux New Zealand Ltd, Suite 4, The Business Connection Office, 2 Kalmia Street, Auckland 1544
Causa
The vitek ms instrument is equipped with a transient recorder (tr) board associated with a specific firmware. this association is used to compute the time of flight of detected ions. some issues have been reported when firmware version 3.13 was installed on systems equipped with the old tr board (ref. 6200350). extensive testing has proven that firmware 3.13 may potentially cause an instrument to freeze or hang during acquisition when installed with the old tr board, with a risk of repeated results for different spots.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
There is a low but existing entrapment risk for a patient being treated of this bed while being under a bmi of 40. small, lightweight patients are at a greater risk of entrapment because their small physical dimensions may enhance their ability to fit into a gap, in whole or in part.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Causa
Ge healthcare has recently become aware of a potential safety issue related to a spect scan procedure used for pulmonary imaging. contact between the detector and the patient's elbow may occur during spect protocols using 450 angular steps in scenarios when a patient's arms are positioned above the head and when dedicated safety armbands are not utilized.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Molnlycke Health Care Ltd, 58 Richard Pearse Drive, Airport Oaks, Mangere, Manukau 2022
Causa
The manufacturer has determined that some units of the basin liners provided in the pack, have small holes in them which could be undetectable in normal use. although the instructions for use of the basin liner itself clearly state that it should not be used to hold fluids or material for invasive use, these holes could, if present potentially compromise the sterile field.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
Causa
There is a software issue which in certain circumstances could result in the unintended deletion of images sent to the intellispace pacs 4.4 before they are available for access by system users., timing of images being received by the pacs varies based on study size or modality configuration. if additional images are received into the study input folder as the input folder is being deleted, those images are not saved and can result in missing images.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
Causa
Manufacturer confirmed that samples containing fluoroscein may show inteference with the advia centaur systems tsh3 ultra, vitamin d and the advia centaur brahms procalcitonin assays.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
Notas adicionales en la data
Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
Causa
The gsi viewer application running on aw/volume share or on the aw server when making a batch and rotating the batch group clockwise 180 degrees (inferior to superior) will display images in the superior orientation. the orientation annotation on the image is correct. the concern is that the patient right/left representation is from the superior view, which could result in a misreading of the image and misdiagnosis.
Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
The ulib database delivered with new devices as well as the ulib database offered between march 25 and sept 24 2013 for download contained an erroneous value for one lens. this could lead to the implantation of a lens with a wrong diopter value., specifically the a0 constant of the lens amo sensar as40e is listed as 0.0 in the internal lens data base of the iol master. the correct value is -2.420.