U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Hip Stems - Product Code LWJ
Causa
Sterility (package integrity) compromised: device package damaged during shipment.
Acción
The recalling firm issued an Important Recall Notice letter dated 10/10/07 to all their customers informing them of the problem and the need to return the product.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEI
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GEJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The company was made aware of instances where the grasper jaws broke during laparoscopic procedures. when the grasper jaws had broken they had done so at the junction of the jaw and the tube. in no instances was it reported to the company that small pieces had broken off of the grasper jaws.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Notas adicionales en la data
Laparoscopic Instrument - Product Code GCJ
Causa
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Acción
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.