• Acerca de la base de datos
  • ¿Cómo usar la IMDD?
  • Descargar la data
  • Preguntas frecuentes
  • Créditos
Vista de la lista Vista de las tarjetas
  • Dispositivo 12
  • Fabricante 10
  • Evento 124969
  • Implante 5
Retiro De Equipo (Recall) de Device Recall S.C.O.R.E.S Units
  • Tipo de evento
    Recall
  • ID del evento
    73157
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1266-2016
  • Fecha de inicio del evento
    2016-01-13
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-09-27
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143282
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Sterilization wrap containers, trays, cassettes & other accessories - Product Code KCT
  • Causa
    Water retention issues. units with filters that remain wet after the sterilization may compromised sterility.
  • Acción
    On January 13, 2016 PMBS, LLC sent letters to their customers informing them of the correction. Per product's instructions for use, the SCORES cabinet is to be run for a minimun of a 4 minute sterilization cycle, followed by a 30 munites dry time followed by a 30 minutes door crack time. not all facilities generate completely dry SCORES filters following these instructions. It is imperative that the filters be checked in the Operation Room prior to useing the surgical instruments sterilised within the cabinet. If the filters are wet, do not use the contents. In the letter is a IFU addendum including instructions for running the equiptment with extended cycles. For more information please contact Ms. Michele Mauzerall, CEO at 908-300-7093.
Retiro De Equipo (Recall) de Device Recall Karma
  • Tipo de evento
    Recall
  • ID del evento
    73159
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1025-2016
  • Fecha de inicio del evento
    2016-01-27
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2016-06-29
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143292
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Wheelchair, mechanical - Product Code IOR
  • Causa
    Karman healthcare, inc. is recalling karma flexx wheelchair because incorrect size screws and corresponding hardware were installed on the front caster fork brackets.
  • Acción
    On 01/27/16 the firm sent out customer notification letters. In the letter customers were informed that the firm was conducting a product update for original purchasers of KM8522. Customers were provided the following procedures for the product update: 1. Completed and signed notification must be returned to Karman Healthcare via fax at (626)-581-2335, or e-mail to edward@karmanhealthcare.com. 2. Following the received frame, customers may install their current wheels, and footrest onto the newly updated frame. Kindly package the used frame into the carton in the same manner the new carton was shipped. 3. A prepaid return shipping label is included with the letter, which customers may use to return the carton; or contact the firm's customer service line at (800)-805-2762 to have a trained representative schedule a pickup at no charge.
Retiro De Equipo (Recall) de Device Recall COOK Stiffened Micropuncture Introducer Set
  • Tipo de evento
    Recall
  • ID del evento
    25109
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0430-03
  • Fecha de inicio del evento
    2002-11-25
  • Fecha de publicación del evento
    2003-02-28
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2003-03-21
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=25208
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Introducer, Catheter - Product Code DYB
  • Causa
    The tip of the stiffened cannula might break during use and remain in the patient.
  • Acción
    Recall letters were sent to each customer on November 26, 2002 advising them of the recall and to return stocks on hand.
Retiro De Equipo (Recall) de Device Recall Flexor CheckFlo Introducer
  • Tipo de evento
    Recall
  • ID del evento
    73163
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0865-2016
  • Fecha de inicio del evento
    2016-01-20
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2016-11-21
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143296
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Introducer, catheter - Product Code DYB
  • Causa
    Incorrect packaging. reports of packages labeled as 6.0fr actually containing 5.0fr devices.
  • Acción
    On 1/20/2016, URGENT: MEDICAL DEVICE RECALL notifications were sent to the hospital Risk Management/Recall Administration via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
Retiro De Equipo (Recall) de Device Recall LEGION Hemi Stepped Tibial Wedge
  • Tipo de evento
    Recall
  • ID del evento
    73164
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0816-2016
  • Fecha de inicio del evento
    2016-01-18
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2016-05-25
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143298
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    The package for catalog item, a legion hemi stepped tibial wedge, 5 mm size 5-6 lt-lat/ rt-mdl, should include both a screw-on hemi stepped tibial wedge and two tibial wedge screws to affix the wedge to the tibial baseplate. a complaint was received indicating that the screws packaged with the wedge were too long.
  • Acción
    The firm notified their direct consignee of the issue on 01/18/2016 by email. The notice requested return of the product.
Retiro De Equipo (Recall) de Device Recall Covidien
  • Tipo de evento
    Recall
  • ID del evento
    73121
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1216-2016
  • Fecha de inicio del evento
    2016-01-22
  • Fecha de publicación del evento
    2016-03-21
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-06-20
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143301
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter, retention type, balloon - Product Code EZL
  • Causa
    Aplicare castile soap towelettes included in the foley trays and mid stream specimen kits are contaminated with bacteria.
  • Acción
    Medtronic/Covidien notfiied customersl via Federal Express on January 22, 2016,, The letter instructs customers to not use the Castile Soap Towelettes contained in the product kits listed on Attachment 1 and discard the affected towelettes. The Castile Soap packet included in these kits does not impact the integrity of the remaining products in the kit. Questions or concerns, contact your Medtronic representative or Customer Service at (800) 882-5878. For questions regarding this recall call 203-492-5000.
Retiro De Equipo (Recall) de Device Recall Covidien
  • Tipo de evento
    Recall
  • ID del evento
    73121
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1217-2016
  • Fecha de inicio del evento
    2016-01-22
  • Fecha de publicación del evento
    2016-03-21
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-06-20
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143302
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter, retention type, balloon - Product Code EZL
  • Causa
    Aplicare castile soap towelettes included in the foley trays and mid stream specimen kits are contaminated with bacteria.
  • Acción
    Medtronic/Covidien notfiied customersl via Federal Express on January 22, 2016,, The letter instructs customers to not use the Castile Soap Towelettes contained in the product kits listed on Attachment 1 and discard the affected towelettes. The Castile Soap packet included in these kits does not impact the integrity of the remaining products in the kit. Questions or concerns, contact your Medtronic representative or Customer Service at (800) 882-5878. For questions regarding this recall call 203-492-5000.
Retiro De Equipo (Recall) de Device Recall Centurion Convenience Kits
  • Tipo de evento
    Recall
  • ID del evento
    73168
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0861-2016
  • Fecha de inicio del evento
    2016-01-19
  • Fecha de publicación del evento
    2016-02-23
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2016-07-19
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143304
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Set, administration, intravascular - Product Code FPA
  • Causa
    Convenience kits contain an in-vision-plus needleless iv connector, code rym5001 that is being recalled by the manufacturer, rymed technologies, llc. a component molding abnormality may cause female luer threads to crack, leading to potential for the mating tubing set to become detached and for the administration to be interrupted. other components within these kits are not affected.
  • Acción
    The firm, RyMed Technologies, LLC, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated 1/4/2016 to the affected customers via certified mail on 1/19/2016 with instructions for identifying and returning the affected product. The recall letter included a description of the reason for the recall, affected product, actions to be taken, and instructions for responding to the formal recall notification. The customers were instructed to discontinue distribution and/or use and return product to: RyMed Technologies, LLC, 6000 W. William Cannon Drive, Building B, Suite 300, Austin, TX 78749; that user facilities change the IV connector within 24 hours when lipids are being administered; if you further distributed any of the lots, contact your accounts, advise them of recall and return their outstanding stock, complete and return enclosed Acknowledgement and Receipt Form ASAP and no later than 10 business days via E-mail: amccutchen@rymedtech.com, Fax: 512-301-7338 or by mail to address above, Attention Recall Return Notification. Please contact Director of Quality Assurance/Regulatory Affairs at (512) 301-7334 Ext.303 or email: ammccutchen@rymedtech.com should you have questions pertaining to recall notification. Questions pertaining to Returning Product/Replacement Product contact Logistics Specialist at (512) 301-7334 Ext.309 or email svara@rymedtech.com.
Retiro De Equipo (Recall) de Device Recall USE PTA 0.014 RX Catheter
  • Tipo de evento
    Recall
  • ID del evento
    73171
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1362-2016
  • Fecha de inicio del evento
    2016-02-11
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-11-30
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143315
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Catheter, angioplasty, peripheral, transluminal - Product Code LIT
  • Causa
    Product may kink during use rendering the product unusable.
  • Acción
    NATEC Medical ltd.sent an Urgent Medical Device Advisory Notice dated March 18, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Warnings: This device should be used only by physicians who are experienced and have a thorough understanding of the clinical and technical aspects of angioplasty. What action do you need to take: 1. Always use a guiding catheter when contralateral access is needed. 2. Always insert cautiously the PTA catheter inside the guiding catheter, and if any resistance is felt, the cause should be determined via fluoroscopy. 3. For contralateral procedure, the USE014OTW has to be preferred to Ebony 014 Rx. NATEC Medical Ltd and our distributor US Endovascular are committed to keeping you and your healthcare professional informed of issues and solutions concerning our products and services. Should you have any questions regarding this or any other matter regarding your USE 014Rx PTA catheter, please do not hesitate to contact our distributor: US Endovascular at (314) 341-4930.
Retiro De Equipo (Recall) de Device Recall NEXGEN COMPLETE KNEE SOLUTION CRUCIATE RETAINING (CR)...
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0831-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143317
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall NEXGEN COMPLETE KNEE SOLUTION LEGACY POSTERIOR STABIL...
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0832-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143318
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall NEXGEN KNEE PROSTHESIS & LEGACY KNEE PROSTHESIS W/CON...
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0833-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143319
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall Ventilator, continuous, lifesupporting; Ventilator, c...
  • Tipo de evento
    Recall
  • ID del evento
    73177
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1065-2016
  • Fecha de inicio del evento
    2016-01-18
  • Fecha de publicación del evento
    2016-03-07
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-26
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143322
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Ventilator, continuous, facility use - Product Code CBK
  • Causa
    Software issue.
  • Acción
    Philips Respironics has determined that affected devices may continue to be used in accordance with the device Instructions for Use and the Field Safety Notification that is being issued to customers. Philips Respironics will be providing a software update for affected devices free of charge. Customers will be notified by mail when the software update is available for download. For further questions please call ( 724 ) 387-7500.
Retiro De Equipo (Recall) de Device Recall AccuLIF Tubing Set
  • Tipo de evento
    Recall
  • ID del evento
    73179
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0852-2016
  • Fecha de inicio del evento
    2015-12-16
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2016-08-01
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143326
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Intervertebral fusion device with bone graft, lumbar - Product Code MAX
  • Causa
    There were two potential interference conditions identified with the way the tubing set attaches to the inserter.
  • Acción
    Stryker Spine sent an Urgent Product Removal letter dated December 17, 2015 to affected customers via Fed Ex priority overnight. The letter identified the affected product, problem and actions to be taken. Customers are asked to follow instructions included in the letter. The customer response form should be completed and sent back to email via Spine-RegulatoryActions@Stryker.com or Fax 855-632-9049. For questions call (201)749-8389.
Retiro De Equipo (Recall) de Device Recall GENDER SOLUTIONS NATURALKNEE FLEX SYSTEM; NEX GEN COM...
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0838-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143345
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall Stryker Spine
  • Tipo de evento
    Recall
  • ID del evento
    73180
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0828-2016
  • Fecha de inicio del evento
    2015-12-18
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2016-08-01
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143327
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Orthopedic manual surgical instrument - Product Code LXH
  • Causa
    Reported jamming of the impaction handle when assembled to the inserter guide at the threads.
  • Acción
    Stryker Spine sent Urgent Product Removal notification letters to branches/agencies on 12/18/2015 via FedEx. The letters provided a description of the affected product, the issue, potential hazards and risk mitigations, and actions to be taken. Customers were asked to examine their inventory and hospital locations to identify the product and reconcile with the affected catalog and lot numbers provided. The Customer Response Form and Product Quantity Tracking Spreadsheet for Health Care Facilities should be completed and returned within five days of receipt of the notice to the e-mail or fax number provided. Questions can be directed to Nabil Riaz - Regulatory Compliance Specialist, Regulatory Compliance & Quality Systems at Nabil.Riaz@Stryker.com.
Retiro De Equipo (Recall) de Device Recall NEXGEN COMPL. KNEE SOL. LEGACY POSTERIOR STABILIZED/C...
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0834-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143331
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall NEXT GENERATION KNEE SYSTEM
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0835-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143333
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall ZIMMER UNICOMPARTMENTAL KNEE SYSTEM
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0836-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143336
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, femorotibial, non-constrained, cemented, metal/polymer - Product Code HSX
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall ZIMMER PATELLOFEMORAL JOINT PROSTHESIS
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0837-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143340
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patello/femoral, semi-constrained, cemented, metal/polymer - Product Code KRR
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall Customed
  • Tipo de evento
    Recall
  • ID del evento
    72992
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1400-2016
  • Fecha de inicio del evento
    2016-01-07
  • Fecha de publicación del evento
    2016-04-14
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-03
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143370
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Syringe, irrigating (non dental) - Product Code KYZ
  • Causa
    Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
  • Acción
    Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016.
Retiro De Equipo (Recall) de Device Recall NATURALKNEE II SYSTEM
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0839-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143350
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall Auto Endo5 Hemolok Automatic Clip Applier
  • Tipo de evento
    Recall
  • ID del evento
    73181
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0886-2016
  • Fecha de inicio del evento
    2016-02-02
  • Fecha de publicación del evento
    2016-02-25
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-08-11
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143353
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Clip, implantable - Product Code FZP
  • Causa
    The clip in the applier may be missing a boss. a boss is the part of the clip that is used to hold the clip within the applier jaws prior to application.
  • Acción
    Teleflex sent an Urgent Medical Device Recall Notification letter dated February 2, 2016 to customers. The letter identified the affected product, problem and actions to be taken. A separate letter was sent to distributors advising them to conduct a sub-recall. Customers were instructed to: 1. If you have affected stock, immediately discontinue use and quarantine any products with the catalog numbers listed in the letter. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1- 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com.This will document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will document receipt of the letter. For questions contact your local sales representative or Customer Service at 1-866-246-6990
Retiro De Equipo (Recall) de Device Recall NEXGEN COMPLETE KNEE SOLUTION LPSFLEX AND KNEE GENDER...
  • Tipo de evento
    Recall
  • ID del evento
    73142
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-0840-2016
  • Fecha de inicio del evento
    2016-01-11
  • Fecha de publicación del evento
    2016-02-22
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-23
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143363
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/ - Product Code MBH
  • Causa
    Ldpe bag containing the implant adheres to the highly polished implant surface.
  • Acción
    Zimmer Biomet sent an Urgent Medical Device Recall dated January 11, 2016, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions or concerns were instructed to contact customer service at 1-800-348-2759 or contact their Zimmer Biomet Sales Representative. For questions regarding this recall call 800-447-5633.
Retiro De Equipo (Recall) de Device Recall Customed
  • Tipo de evento
    Recall
  • ID del evento
    72992
  • Clase de Riesgo del Evento
    Class 2
  • Número del evento
    Z-1398-2016
  • Fecha de inicio del evento
    2016-01-07
  • Fecha de publicación del evento
    2016-04-14
  • Estado del evento
    Terminated
  • País del evento
    United States
  • Fecha de finalización del evento
    2017-05-03
  • Fuente del evento
    USFDA
  • URL de la fuente del evento
    https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=143367
  • Notas / Alertas
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • Notas adicionales en la data
    Suture removal kit - Product Code MCZ
  • Causa
    Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
  • Acción
    Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016.
  • « First
  • ‹ Prev
  • …
  • 5
  • 6
  • 7
  • 8
  • 9
  • …
  • Next ›
  • Last »

Acerca de la base de datos

Explore más de 120,000 registros de retiros, alertas y notificaciones de seguridad de dispositivos médicos y sus conexiones con los fabricantes.

  • Preguntas frecuentes
  • Acerca de la base de datos
  • Contáctenos
  • Créditos

Historias en su correo

¿Trabaja en la industria médica? ¿O tiene experiencia con algún dispositivo médico? Nuestra reportería no ha terminado. Queremos oír de usted.

¡Cuéntanos tu historia!

Aviso

Los dispositivos médicos ayudan con el diagnóstico, la prevención y el tratamiento de muchas lesiones y enfermedades. A través de la International Medical Devices Database no estamos sugiriendo que compañías u otras entidades mencionadas en la base de datos hayan sido parte de una conducta ilegal o hayan actuado de manera impropia. Un mismo dispositivo médico puede tener distintos nombres en diferentes países. Esta base de datos no busca proporcionar asesoría médica. Los pacientes deben consultar con sus médicos para determinar si la data contiene información relevante y si la misma tiene implicaciones médicas para ellos.